Designation: F 2338 – 05
Standard Test Method for
Nondestructive Detection of Leaks in Packages by Vacuum
Decay Method1
This standard is issued under the fixed designation F 2338; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 Test Packages—Packages that can be nondestructively
evaluated by this test method include:
1.1.1 Rigid and semi-rigid non-lidded trays.
1.1.2 Trays or cups sealed with porous barrier lidding
material.
1.1.3 Rigid, nonporous packages.
1.1.4 Flexible, nonporous packages (see 1.2.4).
1.2 Leaks Detected—This test method is capable of detect-
ing package leaks using an absolute or differential pressure
transducer leak detector. The sensitivity of a test is a function
of the sensitivity of the transducer, the package design, the
design of the package test fixture, and critical test parameters
of time and pressure. Types and sizes of leaks that may be
detected for various package systems, as well as test sensitivi-
ties are described below. These data are based on precision and
bias confirmation studies.
1.2.1 Trays or Cups (Non-lidded)—Hole or crack defects in
the wall of the tray/cup of at least 50 µm in diameter can be
detected at a Target Vacuum of 4·104 Pa (400 mbar) using an
absolute pressure transducer test instrument.
1.2.2 Trays Sealed with Porous Barrier Lidding Material—
Hole or crack defects in the wall of the tray/cup of at least 100
µm in diameter can be detected. Channel defects in the seal
area (made using wires of 125 µm in diameter) can be detected.
Severe seal bonding defects in both continuous adhesive and
dot matrix adhesive package systems can be detected. Slightly
incomplete dot matrix adhesive bonding defects can also be
detected. All porous barrier lidding material packages were
tested at a Target Vacuum of 4·104 Pa (400 mbar) using an
absolute pressure transducer test instrument. Using a calibrated
volumetric airflow meter, the sensitivity of the test for porous
lidded packages is shown to be approximately 10-2 Pa·m3·s-1.
1.2.3 Rigid, Nonporous Packages—Hole defects of at least
5 µm in diameter can be detected. All rigid, nonporous
packages were tested at a target vacuum of 5·104 Pa (500 mbar)
using a differential pressure transducer test instrument. Using a
calibrated volumetric airflow meter, the sensitivity of the test
for rigid, nonporous packages is shown to be approximately
10-4 Pa·m3·s-1.
1.2.4 Flexible, Nonporous Packages—Such packages may
also be tested by the vacuum decay method using either an
absolute or differential pressure tranducer test instrument. The
instrument should be selected based on the leak test sensitivity
desired. Sensitivity data for flexible packages were not in-
cluded in the precision and bias studies, although the use of
vacuum decay for testing such packages is well known.
1.3 Test Results—The test results are qualitative (Accept/
Reject). Acceptance criteria for test results are established from
quantitative baseline vacuum decay measurements obtained
from control, non-leaking packages.
1.4 Standard Value Units—The values used in this test
method are stated in SI units and are to be regarded as standard
units. Values in parentheses are for information only.
1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards: 2
D 996 Terminology of Packaging and Distribution Environ-
ments
E 691 Practice for Conducting an Interlaboratory Study to
Determine the Precision of a Test Method
F 17 Terminology Relating to Flexible Barrier Materials
F 1327 Terminology Relating to Barrier Materials for Medi-
cal Packaging
3. Terminology
3.1 Definitions—For definitions used in this test method,
see Terminologies D 996, F 17, and F 1327.
1 This test method is under the jurisdiction of ASTM Committee F02 on Flexible
Barrier Materials and is the direct responsibility of Subcommittee F02.40 on
Package Integrity.
Current edition approved April 1, 2005. Published April 2005. Originally
approved in 2003. Last previous edition approved in 2004 as F 2338 – 04.
2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
1
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Oliver Pace (); Fri May 5 08:38:39 EDT 2006
3.2 Definitions of Terms Specific to This Standard:
3.2.1 baseline vacuum decay, n—the extent of vacuum
change within the test chamber over time demonstrated by a
control, non-leaking package.
3.2.2 control, non-leaking packages, n—packages without
defects and properly sealed or closed according to manufac-
turer’s specifications.
3.2.3 flexible, nonporous packages, n—packages that sig-
nificantly deflect when under vacuum, and are constructed of
malleable, nonporous materials. Examples include pouches or
bags made of polymeric, foil, or laminate films.
3.2.4 rigid, nonporous packages, n—packages that do not
significantly deflect under vacuum and are constructed of solid,
nonporous materials. For example, plastic bottles with screw-
thread or snap-on closures are rigid, nonporous packages.
3.2.5 semi-rigid trays or cups, n—trays made of material
that retain shape upon deflection. For example, thermoformed
PETE or PETG trays are considered semi-rigid trays.
3.2.6 spotty or mottled seals, n—an incomplete adhesive
bond made between a package tray or cup and porous lidding
material that can be visibly identified by a distinctive pattern of
dots, spotting or mottling on the tray sealing surface after the
lid is removed.
3.2.7 volumetric airflow meter, n—a calibration tool that
can be used to provide an artificial leak of known volumetric
airflow rate into the test chamber for verification of instrument
sensitivity. Airflow meters should be calibrated to NIST
standards. The operational range of the meter should reflect the
desired limit of sensitivity for the intended leak test.
3.3 Definitions of Test Cycle and Critical Parameters
Terms—For terms and abbreviations relating to the test cycle
and the critical parameters for establishing accept/reject limits,
see Annex A1.
4. Summary of Test Method
4.1 The test package is placed in a test chamber to which
vacuum is applied. The chamber is then isolated from the
vacuum source and an absolute or differential vacuum trans-
ducer is used to monitor the test chamber for both the level of
vacuum, as well as the change in vacuum over time. Vacuum
decay, or rise in chamber pressure, is a result of package
headspace gas being drawn out of the package through any
leaks present, plus background noise. Leak detection requires
vacuum decay in excess of the background noise level.
Background noise vacuum decay may result from package
expansion when exposed to vacuum (flexible or semi-rigid
packages), or from residual gases inherent in the test chamber
or test system lines.
4.2 Porous barrier lidded tray or cup packages are tested for
leaks located in the tray or cup, and at the lidding material/tray
seal junction. Leaks in the porous lidding material itself cannot
be detected. When testing such packages, steps are taken to
physically mask or block the porous barrier surface to prevent
the migration of package gas through the porous lid. These
steps may require some sample preparation, depending on the
masking approach required, but must be nondestructive and
noninvasive. Vacuum decay from porous barrier lidded pack-
ages may potentially include background noise from gas
trapped between the lidding material and the masking surface,
or from transverse gas flow through the porous barrier material
itself at the lid/tray seal junction.
4.3 The sensitivity of a vacuum decay leak test is a function
of several factors. Smaller leaks can be detected with more
sensitive pressure transducers, and with longer test times. Also,
pressure changes can be more readily detected with smaller
void volumes between the test package and the test chamber,
and with smaller test system line volumes. Steps to reduce
background noise can also improve sensitivity. For example,
for porous barrier lidded packages, more effective masking
techniques will minimize background noise.
NOTE 1—Further information on the “Leak Test Theory” may be found
in Annex A1.
5. Significance and Use
5.1 Leaks in medical device, pharmaceutical and food
packages may result in the ingress of unwanted gases (most
commonly oxygen), harmful microbiological or particulate
contaminants. Package leaks may appear as imperfections in
the package components themselves or at the seal juncture
between mated components. The ability to detect leaks is
necessary to ensure consistency and integrity of packages.
5.2 After initial set-up and calibration, the operations of
individual tests may be semi-automatic, automatic or manual.
The test method permits the non-destructive detection of leaks
not visibly detectable. The test method does not require the
introduction of any extraneous materials or substances, such as
dyes or gases. However, it is important to physically mask or
block off any porous barrier surface of the package during the
test to prevent a rapid loss of chamber vacuum resulting
primarily from gas migration through the porous surface. Leak
detection is based solely on the ability to detect the change in
pressure inside the test chamber as a result of air egress from
the properly masked package when challenged with vacuum
conditions.
5.3 This test is a useful research tool for optimization of
package sealing parameters and for comparative evaluation of
various packages and materials. This test method is also
applicable to production settings as it is rapid, non-invasive
and non-destructive, making it useful for either 100 % on-line
testing or to perform tests on a statistical sampling from the
production operation.
5.4 Leak test results that exceed the permissible limits for
the vacuum decay test are indicated by audible or visual signal
responses, or both.
6. Apparatus
6.1 Vacuum Decay Leak Detection Apparatus—All vacuum
decay test systems include a test chamber with a lower
compartment (lower tooling) designed to nest the test package,
and an upper lid (top tooling) for closing the test chamber. Fig.
1 illustrates a test chamber designed for testing packages with
porous barrier lidding material. The test fixture upper lid
consists of a flexible bladder to mask the package’s porous
barrier during the test cycle. Fig. 2 illustrates a test chamber
designed for testing rigid, nonporous packages. In this case,
there is no flexible bladder. For both test chamber designs, the
test chamber is connected to the vacuum decay test system.
This system includes a vacuum source for establishing vacuum
F 2338 – 05
2
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Oliver Pace (); Fri May 5 08:38:39 EDT 2006
within the chamber at the beginning of the test cycle, and an
absolute or differential pressure transducer for monitoring the
level of vacuum as well as the pressure change as a function of
time during the test cycle. A calibrated volumetric airflow
meter may be placed in-line with the test system for verifying
the sensitivity of a leak test.
6.2 Tray Nest or Lower Tooling—The bottom half of the test
chamber is dimensionally designed to closely nest the test
package, while still allowing for easy gas flow around the test
package. Without ready gas flow around the package, leakage
sites can be blocked. Conversely, the larger the gap between
the test chamber and the test package, the less sensitive the leak
test, as vacuum decay from package leakage will be minimized
in a larger net test chamber volume.
6.3 Upper Lid or Upper Tooling—The upper lid is designed
to tightly seal the closed test chamber during the vacuum cycle.
6.4 Mask or Block—The porous barrier lidding material of
packages must be masked or blocked during testing to mini-
mize egress of air from the package through the lidding.
Various masking techniques may be used, including a test
chamber designed with a flexible bladder in the upper tooling
(refer to Fig. 1).
6.5 Volumetric Airflow Meter—An adjustable volumetric
airflow meter is placed in-line with the test chamber to
introduce an artificial leak of variable size. It is recommended
that an airflow meter be used to verify the sensitivity of the leak
test parameters.
NOTE 2—Refer to Annex A2 for further information about the use of a
volumetric airflow meter for verifying leak test sensitivity.
7. Hazards
7.1 As the test chamber is closed, it may present pinch-point
hazards.
8. Preparation of Apparatus
8.1 The test apparatus must be started, warmed-up, and
made ready according to the manufacturer’s specifications.
Utilities required for instrument operation include electrical
power and a supply of dry, non-lubricated compressed air,
according to manufacturer’s specifications.
9. Calibration and Standardization
9.1 Before test measurements are made, the apparatus must
be calibrated. The pressure transducers, the vacuum source
pressure gage, and the adjustable volumetric airflow meter
must all be calibrated according to the manufacturer’s recom-
mended procedures and maintenance schedule.
9.2 Critical test parameter settings must be established for
each package/test fixture combination. Parameters will vary
based on the test package geometry and any porous barrier
surface’s inherent porosity.
NOTE 3—Refer to Section 4 and Annex A1 for a description of critical
test parameters.
FIG. 1 Schematic of Fixture and Porous Barrier Lidded Test Package
F 2338 – 05
3
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Oliver Pace (); Fri May 5 08:38:39 EDT 2006
9.3 A sample population of control non-leaking packages
must be used for selecting and optimizing critical test param-
eters. Control packages are to be made from the same materials
and according to the same design as the test units.
NOTE 4—Refer to Annex A2 for information on critical test parameter
selection.
9.4 After critical test parameters have been selected, qualify
the ability of the test to reliably differentiate between known
non-leaking and defective packages.
9.5 Determine the sensitivity of the test using control
non-leaking test packages and a calibrated volumetric airflow
meter.
NOTE 5—Refer to Annex A2 for information about test sensitivity
verification procedures.
9.6 Test qualification (see 9.4) and test sensitivity verifica-
tion (see 9.5) are to be conducted frequently, typically at least
one or more times a day, preferably at the beginning of every
shift.
10. Procedure
10.1 Select and install the appropriately sized test chamber
for the package to be tested. Make any necessary adjustments
to the chamber to ensure a sufficiently tight seal of the chamber
lid (upper tooling) to the lower chamber package nest (lower
tooling) when the test chamber is in the closed position.
FIG. 2 Schematic of Fixture and Rigid, Nonporous Test Package
F 2338 – 05
4
Copyright by ASTM Int'l (all rights reserved);
Reproduction authorized per License Agreement with Oliver Pace (); Fri May 5 08:38:39 EDT 2006
10.2 Verify the pressure level available at the supply source.
Check the functionality of the vacuum source.
10.3 Program the test instrument with all necessary test
parameters and accept/reject criteria.
10.4 Place the assembled package into the lower tooling
nest and close the test chamber. Take appropriate steps to mask
or block any porous barrier surface of the package.
NOTE 6—Inspect and clean the masking or blocking surface according
to a regularly established routine according to the instrument manufac-
turer’s recommended procedures to ensure effective masking of the porous
barrier surface.
10.5 Start the test.
10.6 Note the pass or fail indicator or other means of
detecting vacuum decay and document results. Identify and set
aside any failed package for further evaluation.
10.7 Select another package and repeat the testing process.
11. Report
11.1 For each package tested, report the values for the
following critical test parameters as well as package test
results:
11.1.1 Reserve Vacuum (VRes),
11.1.2 Target or Test Vacuum (Vac),
11.1.3 Reference Vacuum (VRef),
11.1.4 Reference Fill Time (TfillRef),
11.1.5 Equalization Time (Tequal),
11.1.6 Vacuum Decay Test Time (Ttest),
11.1.7 Reference Vacuum Decay Rate Accept/Reject Limit
(dP/dt Ref), and
11.1.8 Accept/Reject Test Results.
NOTE 7—Refer to Annex A1 for definitions of critical test parameters.
12. Precision and Bias
12.1 Nonlidded and Porous Barrier Lidded Trays—An
interlaboratory study was run in accordance with Practice
E 691 using an absolute pressure decay instrument.3 Three
laboratories ran the study, each using a separate instrument.
Each laboratory performed three replicate tests on each test
sample. Test sample populations consisted of non-lidded semi-
rigid (PETE) thermoformed trays, and trays sealed with vari-
ous porous barrier lidding materials. The same test samples
were tested at each laboratory.4 Test results are qualitative in
nature (Pass or Fail). Operators selected test critical parameters
for each sample population; therefore test results reflect opera-
tor, laboratory and instrument variability.
12.1.1 Nonlidded Trays—As summarized in Table 1, two
populations of non-lidded trays representing two tray sizes
were tested. Defective samples contained a single hole in the
tray wall of either 50 µm or 100 µm in diameter. Two of the five
larger trays, each with a 50 µm hole, repeatedly failed to be
detected at more than one test site, while the other three trays
were consistently identified as leaking. At the completion of
the study the two suspect trays were independently reexamined
for the presence and size of the holes. It was determined that
the holes could no longer be located and it was hypothesized
that they had become clogged. Eliminating the two suspect
trays, results demonstrate the ability of the test method to
identify defective trays with holes $50 µm, when using a
Target Vacuum (Vac) of 4·104 Pa (400 mbar).
12.1.2 Porous Barrier Lidded Trays—As per the results
outlined in Table 2, two populations of porous barrier lidded
tray packages were tested, representing two package sizes, all
sealed with one type of coated porous barrier lidding material.
Defective samples included packages with a single hole in the
tray wall (50 µm or 100 µm in diameter), and packages with a
single seal channel defect created using a wire of either 75 µm,
100 µm, or 125 µm in diameter (0.003, 0.004, and 0.005 in.,
respectively). An independent laboratory microscopically veri-
fied tray hole sizes, however seal channel sizes could not be
reliably verified. Results demonstrate the ability of the test
method to identify defective packages sealed with porous
barrier lidding material. Defects consistently identified include
tray holes of at least 100 µm in diameter, and channel defects
created using a 125 µm wire, when using a Target Vacuum of
4·104 Pa (400 mbar).
12.1.3 Porous Barrier Lidded Trays with Various Adhesive
Bonding Systems—Table 3 documents test results using two
populations of tray packages wi
本文档为【ASTMF2338[1]】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。