Technical Information
Oxymax H COS21
Sensor with long-term stability for frequent sterilization and autoclaving
Application
• Process control in enzyme production
• Control of culture growth
• Biotechnological production
• Food industry
• General process applications
Your benefits
• Sensor version suitable for the pharmaceutical industry:
– Stainless steel 1.4435 (AISI 316L)
– Sterilizable and autoclavable
• Application-specific versions:
– Sensor for standard applications
• Versatile usage:
– Standard process connection Pg 13.5
– Installation in standard pH assemblies possible
• Short response time: t98 < 60 s
• Integrated temperature sensor
Oxymax H COS21
2 Endress+Hauser
Function and system design
Measuring principle The oxygen molecules diffused through the membrane are reduced to hydroxide ions (OH-) at the cathode.
Silver is oxidized to silver ions (Ag+) at the anode (this forms a silver halogenide layer).
A current flows due to the electron donation at the cathode and the electron acceptance at the anode. Under
constant conditions, this flow is proportional to the oxygen content of the medium.
This current is converted in the transmitter and indicated on the display as an oxygen concentration in mg/l,
as a saturation index in % SAT or as an oxygen partial pressure in hPa.
Measuring system A complete measuring system comprises:
• Oxygen sensor COS21
• Transmitter, e.g. Liquisys M COM223/253 F
• Special measuring cable COK21
• Assembly, e.g. retractable assembly CPA475
a0006653
Fig. 1: Measuring system (example)
1 Oxygen sensor COS21
2 Retractable assembly CPA475
3 Special measuring cable COK21
4 Transmitter Liquisys M COM253 F
Input
Measured variable dissolved oxygen [mg/l / % SAT / hPa]
Measuring range 0,01 to 20 mg/l (ppm)
0 to 200% SAT
1
2
3
4
CAL REL
REL1
REL1
ALARM
REL2
REL2E
Oxymax H COS21
Endress+Hauser 3
Wiring
Electrical connection The electrical connection between the sensor and the transmitter uses the multi-core special measuring cable
COK21.
Performance characteristics
Response time From air to nitrogen at 25 °C (77 °F)
• t90 : < 30 s
• t98 : < 60 s
Reference operating
conditions
Minimum flow rate 0.03 m/s (0.1 ft/s)
Zero current < 0.1 % of the current at air
Maximum measured error ±1 % of measured value1)
Repeatability ±0.1 % of measuring range end
Long-term drift
Influence of medium
temperature
approx. 3.1% per K (compensated by transmitter)
Influence of medium pressure Pressure compensation not necessary
Polarization time < 60 minutes
Polarization voltage –670 ± 50 mV
Oxygen intrinsic consumption approx. 20 ng/h in air at 25 °C (77 °F)
a0005583-en
Special measuring cable COK21
PK
GN / YE
BN
YE
BU
S
12
90
91
11
Assignment Terminal
External shield
Cathode
Anode
NTC temp. sensor
Active internal shield
(NTC temp. sensor)
Reference temperature: 25 °C (77 °F)
Reference pressure: 1013 hPa (15 psi)
1) In accordance with IEC 746-1 at nominal operating conditions
Zero-point drift: < 0.1 % per week at 30 °C (86 °F) and under constant conditions
Measuring range drift: < 0.1 % per week at 30 °C (86 °F) and under constant conditions
Oxymax H COS21
4 Endress+Hauser
Installation
Angle of installation
Environment
Ambient temperature range 0 to +130 °C (32 to 270 °F)
Storage temperature –10 to +60 °C (10 to 140 °F) at 95% relative air humidity, not condensing
" Caution! Danger of drying out
Only store the sensor with the electrode protection cap (filled with 0.02 n NaOH).
Process
Process temperature 0 to 130 °C (32 to 270 °F)
! Note! Please note the temperature compensation range of the transmitter used.
Process pressure 0 to 4 bar (0 to 58 psi)
Sterilization When autoclaving, tightly place protection cap on plug-in head. No plug-in head protection is required when
the device is sterilised in-situ.
! Note! If water has penetrated the plug-in head, dry it with warm or compressed air. This will prevent corrosion or
contact problems.
a0006654-en
Permitted installation angle
10° 10° Not permissible!Not permissible!
Permissible angle of installation
Oxymax H COS21
Endress+Hauser 5
Mechanical construction
Design, dimensions
Weight 0.2 kg (0.44 lbs)
Material
Process connection Thread Pg 13.5
Temperature sensor NTC 22 kΩ
Electrolyte Alkaline electrolyte
Ordering information
Product structure
a0006656
Exploded view
1 Threaded connection Pg 13.5
2, 3 Seals
4 Cathode
5 Protective cap
6 Membrane
7 Membrane cap
8 Anode
9 Plug-in head
a0006655
Dimensions
1
3
4
5
6
7
8
9
2
Sensor shaft:
Membrane:
Electrode combination:
Sealing ring:
Stainless steel 1.4435 (AISI 316L)
Optiflow®
Silver / Platinum
Viton®
Immersion depth
1 120 mm (4.72 in)
2 225 mm (8.86 in)
3 350 mm (13.8 in)
5 120 mm (4.72 in) + EN10204-3.1
6 225 mm (8.86 in) + EN10204-3.1
7 350 mm (13.8 in) + EN10204-3.1
Head
K Process Pg 13.5, four-pole plug-in head
Application
0 Hygienic version
COS21- Complete order code
Oxymax W COS21 1
Oxymax W COS21 2
Oxymax W COS21 3
Oxymax W COS21 4
Oxymax W COS21 5
Oxymax W COS21 6
本文档为【德国E+H分析仪表:溶解氧传感器 Oxymax H COS21】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。