Designation: D 6866 – 04
Standard Test Methods for
Determining the Biobased Content of Natural Range
Materials Using Radiocarbon and Isotope Ratio Mass
Spectrometry Analysis1
This standard is issued under the fixed designation D 6866; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 These test methods do not address environmental im-
pact, product performance and functionality, determination of
geographical origin, or assignment of required amounts of
biobased carbon necessary for compliance with federal laws.
1.2 These test methods are applicable to any product con-
taining carbon-based components that can be combusted in the
presence of oxygen to produce carbon dioxide (CO2) gas.
1.3 These test methods make no attempt to teach the basic
principles of the instrumentation used although minimum
requirements for instrument selection are referenced in the
References section. However, the preparation of samples for
the above methods is described. No details of instrument
operation are included here. These are best obtained from the
manufacturer of the specific instrument in use.
1.4 Currently, there are no ISO test methods that are
equivalent to the test methods outlined in this standard.
1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards: 2
D 883 Terminology Relating to Plastics
3. Terminology
3.1 The definitions of terms used in this test method are
referenced in order that the practitioner may require further
information regarding the practice of the art of isotope analysis
and to facilitate performance of the method.
3.2 Terminology D 883 should be referenced for terminol-
ogy relating to plastics. Although an attempt to list terms in a
logical manner (alphabetically) will be made as some terms
require definition of other terms to make sense.
3.3 Definitions:
3.3.1 dpm—disintegrations per minute. This is the quantity
of radioactivity. The measure dpm is derived from cpm or
counts per minute (dpm = cpm − bkgd / counting efficiency).
There are 2.2 by 106 dpm / uCi (11,13).3
3.3.2 dps—disintegrations per second (rather than minute as
above) (11,13).
3.3.3 scintillation—the sum of all photons produced by a
radioactive decay event. Counters used to measure this as
described in this method are Liquid Scintillation Counters
(LSC) Bq (11,13).
3.3.4 specific activity (SA)—refers to the quantity of radio-
activity per mass unit of product, that is, dpmh % (11,13).
3.3.5 automated effıciency control (AEC)—a method used
by scintillation counters to compensate for the effect of
quenching on the sample spectrum (11).
3.3.6 AMS facility—a facility performing Accelerator Mass
Spectrometry.
3.3.7 accelerator mass spectrometry (AMS)—an ultra-
sensitive technique for measuring naturally occurring radio
nuclides, in which sample atoms are ionized, accelerated to
high energies, separated on basis of momentum, charge, and
mass, and individually counted in Faraday collectors. This high
energy separation is extremely effective in filtering out isobaric
interferences, such that AMS may be used to measure accu-
rately the 14C abundance to a level of 1 in 1015. At these levels,
uncertainties are based on counting statistics through the
Poisson distribution (7,8).
3.3.8 background radiation—the radiation in the natural
environment; including cosmic radiation and radionuclides
present in the local environment, for example, materials of
construction, metals, glass, concrete (1,3,6,7,11-15).
3.3.9 coincidence circuit—a portion of the electronic analy-
sis system of a Liquid Scintillation Counter which acts to reject
pulses which are not received from the two Photomultiplier
Tubes (that count the photons) within a given period of time
1 These test methods are under the jurisdiction of ASTM Committee D20 on
Plastics and are the direct responsibility of Subcommittee D20.96 on Environmen-
tally Degradable Plastics and Biobased Products.
Current edition approved February 1, 2004. Published March 2004.
2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
3 The boldface numbers in parentheses refer to the list of references at the end of
this standard.
1
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
and are necessary to rule out background interference and
required for any LSC used in this method (6,11,13).
3.3.10 coincidence threshold—the minimum decay energy
required for a Liquid Scintillation Counter to detect a radioac-
tive event. The ability to set that threshold is a requirement of
any LSC used in this method (11,13).
3.3.11 contemporary carbon—a direct indication of the
relative contributions of fossil carbon and “living” biospheric
carbon can be expressed as the fraction (or percentage) of
contemporary carbon, symbol fC. This is derived from fM
through the use of the observed input function for
atmospheric 14C over recent decades, representing the com-
bined effects of fossil dilution of 14C (minor) and nuclear
testing enhancement (major). The relation between fC and fM is
necessarily a function of time. By 1985, when the particulate
sampling discussed in the cited reference the fM ratio had
decreased to ca. 1.2 (7,8).
3.3.12 chemical quenching—a reduction in the scintillation
intensity (a significant interference with this method) seen by
the Photomultiplier Tubes (PMT, pmt) due to the materials
present in the scintillation solution that interfere with the
processes leading to the production of light. The result is fewer
photons counted and a lower efficiency (3,6,13).
3.3.13 chi-square test—a statistical tool used in radioactive
counting in order to compare the observed variations in repeat
counts of a radioactive sample with the variation predicted by
statistical theory. This determines whether two different distri-
butions of photon measurements originate from the same
photonic events. LSC instruments used in this measurement
should include this capability (11,13,22).
3.3.14 cocktail—the solution in which samples are placed
for measurement in a LSC. Solvents and Scintillators (chemi-
cals that absorbs decay energy transferred from the solvent and
emits light (photons) proportional in intensity to the decay
energy) (3,6,11,13).
3.3.15 decay (radioactive)—the spontaneous transforma-
tion of one nuclide into a different nuclide or into a different
energy state of the same nuclide. The process results in a
decrease, with time, of the number of original radioactive
atoms in a sample, according to the half-life of the radionuclide
(7,11,13).
3.3.16 discriminator—an electronic circuit which distin-
guishes signal pulses according to their pulse height or energy;
used to exclude extraneous radiation, background radiation,
and extraneous noise from the desired signal (11,13,14,26).
3.3.17 effıciency—the ratio of measured observations or
counts compared to the number of decay events which oc-
curred during the measurement time; expressed as a percentage
(11,13).
3.3.18 external standard—a radioactive source placed adja-
cent to the liquid sample in to produce scintillations in the
sample for the purpose of monitoring the sample’s level of
quenching. Required with Method (A) (11,13).
3.3.19 figure of merit—a term applied to a numerical value
used to characterize the performance of a system. In liquid
scintillation counting, specific formulas have been derived for
quantitatively comparing certain aspects of instrument and
cocktail performance and the term is frequently used to
compare efficiency and background measures (11,13,16).
3.3.20 fluorescence—the emission of light resulting from
the absorption of incident radiation and persisting only as long
as the stimulation radiation is continued (11,13,20).
3.3.21 fossil carbon—carbon that contains essentially no
radiocarbon because its age is very much greater than the 5730
year half-life of 14C (7,8).
3.3.22 half-life—the time in which one half the atoms of a
particular radioactive substance disintegrate to another nuclear
form. The half-life of 14C is 5730 years (7,11,20).
3.3.23 intensity—the amount of energy, the number of
photons, or the numbers of particles of any radiation incident
upon a unit area per unit time (11,13).
3.3.24 internal standard—a known amount of radioactivity
which is added to a sample in order to determine the counting
efficiency of that sample. The radionuclide used must be the
same as that in the sample to be measured, the cocktail should
be the same as the sample, the activity of the Internal Standard,
and the Internal Standard must be of certified activity (11,13).
3.3.25 modern carbon—explicitly, 0.95 times the specific
activity of SRM 4990b (the original oxalic acid radiocarbon
standard), normalized to d13C = −19 % (Currie, et al., 1989).
Functionally, the fraction of modern carbon = (1/0.95) where
the unit 1 is defined as the concentration of 14C contempora-
neous with 1950 wood (that is, pre-atmospheric nuclear test-
ing) and 0.95 are used to correct for the post 1950 bomb 14C
injection in to the atmosphere (8).
3.3.26 noise pulse—a spurious signal arising from the
electronics and electrical supply of the instrument
(11,13,17,24).
3.3.27 phase contact—the degree of contact between two
phases of heterogeneous samples. In liquid scintillation count-
ing, better phase contact usually means higher counting effi-
ciency (11,13).
3.3.28 photomultiplier tube (PMT, pmt)—the device in the
LSC that counts the photons of light simultaneously at two
separate detectors (24,26).
3.3.29 pulse—the electrical signal resulting when photons
are detected by the Photomultiplier tubes (11,13,14,26).
3.3.30 pulse height analyzer (PHA)—an electronic circuit
which sorts and records pulses according to height or voltage
(11,13,14,26).
3.3.31 pulse index—the number of afterpulses following a
detected coincidence pulse (used in three dimensional or pulse
height discrimination) to compensate for the background of a
liquid scintillation counter performing (11,14,24,26).
3.3.32 quenching—any material that interferes with the
accurate conversion of decay energy to photons captured by the
PMT of the LSC (1,3,6,11,12,13,16).
3.3.33 region—regions of interest, also called window
and/or channel in regard to liquid scintillation counters. Refers
to an energy level or subset specific to a particular isotope
(3,11,14,17,24).
3.3.34 scintillation reagent—chemicals that absorbs decay
energy transferred from the solvent and emits light (photons)
proportional in intensity to the decay energy (3,11,24).
D 6866 – 04
2
3.3.35 solvent—in scintillation reagent, chemical(s) which
act as both a vehicle for dissolving the sample and scintillator
and the location of the initial kinetic energy transfer from the
decay products to the scintillator; that is, into excitation energy
that can be converted by the scintillator into photons
(3.11,13,24).
3.3.36 standard count conditions (STDCT)—LSC condi-
tions under which reference standards and samples are
counted.
3.3.37 three dimensional spectrum analysis—the analysis of
the pulse energy distribution in function of energy, counts per
energy, and pulse index. It allows for auto-optimization of a
liquid scintillation analyzer allowing maximum performance.
Although different Manufacturers of LSC instruments call
Three Dimensional Analysis by different names, the actual
function is a necessary part of this method (11,13,14).
3.3.38 true beta event—an actual count which represents
atomic decay rather than spurious interference (9,10).
3.3.39 flexible tube cracker—the apparatus in which the
sample tube (Break Seal Tube) is placed (4,5,9,10).
3.3.40 break seal tube—the sample tube within which the
sample, copper oxide, and silver wire is placed.
4. Significance and Use
4.1 Presidential (Executive) Orders 13101, 13123, 13134,
Public Laws (106-224, 107.117, AG ACT 2003 and other
Legislative Actions all require Federal Agencies to develop
procedures to identify, encourage and produce products de-
rived from biobased, renewable, sustainable and low environ-
mental impact resources so as to promote the Market Devel-
opment Infrastructure necessary to induce greater use of such
resources in commercial, non food, products.
4.2 Test Method A utilizes Liquid Scintillation Counting
(LSC) radiocarbon (14C) techniques to quantify the biobased
content of a given product with maximum total error of 15 %
count; associated with sample preparation and actual counting.
4.3 Test Method B utilizes Accelerator Mass Spectrometry
(AMS) and Isotope Ratio Mass Spectrometry (IRMS) tech-
niques to quantify the biobased content of a given product with
possible uncertainties of 1 to 2 % and 0.1 to 0.5 %, respec-
tively. Sample preparation methods are identical to Method A,
9.2–9.5. Method B diverges after 9.5 and rather than LSC
analysis the sample CO2 remains within the vacuum manifold
and is distilled, quantified in a calibrated volume, transferred to
a quartz tube, torch sealed. Details are given in 12.7-12.10. The
stored CO2 is then delivered to an AMS facility for final
processing and analysis.
4.4 Although Method A is less sensitive than that of using
AMS/IRMS, it has the two distinct advantages: (1) lower costs
per evaluation, and (2) much higher instrument availability
worldwide. LSC is the preferred method to be used when
precise dating (2 % uncertainty or less) is not required and
when sample size is not an issue. Indeed, LSC is the most
widely used measurement for 14C determination. Method B
will be used primarily in extraordinary situations such as when
the authenticities of the LSC radiocarbon results are in dispute
or when sample size is greatly restricted or costly per mass of
sample.
4.5 The test methods described here directly discriminate
between product carbon resulting from contemporary carbon
input and that derived from fossil-based input. A measurement
of a product’s 14C/12C content is determined relative to the
modern carbon-based oxalic acid radiocarbon Standard Refer-
ence Material (SRM) 4990c, (referred to as HOxII). It is
compositionally related directly to the original oxalic acid
radiocarbon standard SRM 4990b (referred to as HOxI), and is
denoted in terms of fM, that is, the sample’s fraction of modern
carbon. (See Terminology, Section 3.)
4.6 Reference standards, available to all laboratories prac-
ticing these methods, must be used properly in order that
traceability to the primary carbon isotope standards are estab-
lished, and that stated uncertainties are valid. The primary
standards are SRM 4990c (oxalic acid) for 14C and RM 8544
(NBS 19 calcite) for 13C. These materials are available for
distribution in North America from The National Institute of
Standards and Technology (NIST), and outside North America
from the International Atomic Energy Agency (IAEA), Vienna,
Austria.
4.7 Acceptable SI unit deviations (tolerance) for the practice
of these methods is 65 % from the stated instructions unless
otherwise noted.
5. Safety
5.1 The specific safety and regulatory requirements associ-
ated with radioactivity, sample preparation, and instrument
operation are not addressed in this standard. It is the respon-
sibility of the user of this standard to establish appropriate
safety and health practices. It is also incumbent on the user to
conform to all the Federal and State regulatory requirements,
especially those that relate to the use of open radioactive
source, in the performance of these methods. Although
Carbon-14 is one of the safest Isotopes to work with, State and
Federal regulations must be followed in the performance of this
method.
5.2 The use of glass and metal, in particular with closed
systems containing oxygen that are subjected to 700°C tem-
peratures pose their own safety concerns and care should be
taken to protect the operators from implosion/explosion of the
glass tube. Strong bases used for CO2 absorption, including
Carbosorbt E+ which is also flammable, are particularly
hazardous and instructions in Material Safety Data Sheets
should be followed with special concern for eye protection.
Radioactive carbon-14 compounds should be handled and
disposed of in accordance with State and Federal Regulations.
METHOD A
6. Detailed Requirements
NOTE 1—Acceptable tolerance levels of 65 % are standard to this
method unless otherwise stated.
6.1 Low Level Liquid Scintillation analyzers with active
shielding.
6.2 Anticoincidence systems such as 2 and 3 Photomulti-
plier Tubes (Multidetector systems).
6.3 Coincidence circuits.
D 6866 – 04
3
6.4 Software that includes thresholds, and statistics; Pulse
rise and shape discrimination, and three Dimensional Spectrum
Analysis.
6.5 Use of External and Internal Standards.
6.6 Optimized counting regions to provide very low back-
ground counts while maintaining counting efficiency greater
than 60 % of samples 0.7 to 1.5 g in pre-cleaned, 20 mL Low
Potassium glass count vials.
6.7 No single Liquid Scintillation Counter is specified for
this Method. However, minimum counting efficiency and
control of background interference is specified. Like all ana-
lytical instruments, LS counters require study as to their
specific components and counting optimization.
6.8 Standardization of sample preparation is required.
6.9 External and internal standards must be used in LSC
operation.
6.10 Standardization and optimization of sample vials;
which must be of low Potassium, glass with PTFE insert tops,
20 mL volume, and pre-cleaned. Note: Plastic vials should not
be used for this method.
6.11 Optimization of Scintillation Cocktail, CO2 trap re-
agents, and additives such as emulsifiers, surfactants, alcohols,
and sample material is required.
6.12 Optimization of reagents shall include sample to re-
agent volume, scintillator to CO2 trap reagent and additive
volumes and compatibility of all reagents.
6.13 Quench curves and counting efficiency and back-
ground optimization should be performed using a reference
standard based on comparison to NIST oxalic acid reference
standards and the same reagents and counting parameters as the
samples.
6.14 Counting efficiency of the sample shall be determined
by adding to a vial a known activity of the same radionuclide
and computing the increase of the sample cpm.
6.15 The internal standard technique for computing count-
ing efficiency should be calculated by the equation:
E 5
~cpmb 2 cmpa!
D (1)
where:
E = counting efficiency,
cpma = standard counts of (STDCT) sample without added
known activity,
cpmb = counts of (STDCT) sample plus added known
activity, and
D = dpm of the internal standard.
Other counting interference concerns that must be addressed
as part of specific instrument calibration and normalization
include luminance, static electricity, random noise, temperature
and humidity variability.
6.16 Alternate regions of interest parameters may be used
based upon testing of twenty, or more, 6 h counts of the same
reference (STDCT) standard that record the raw data and
spectrum for KEV regions of interest 4 through 96. Optimal
energy windows counting should be determined by comparison
of E2/bkg values to obtain the highest count efficiency and the
lowest background and other interference. Counting efficiency
of
本文档为【ASTM D6866 – 04】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。