Designation: D 882 – 09
Standard Test Method for
Tensile Properties of Thin Plastic Sheeting1
This standard is issued under the fixed designation D 882; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
These test methods have been approved for use by agencies of the Department of Defense to replace Method 1013 of Federal Test
Method Standard 406.
1. Scope*
1.1 This test method covers the determination of tensile
properties of plastics in the form of thin sheeting, including
film (less than 1.0 mm (0.04 in.) in thickness).
NOTE 1—Film has been arbitrarily defined as sheeting having nominal
thickness not greater than 0.25 mm (0.010 in.).
NOTE 2—Tensile properties of plastics 1.0 mm (0.04 in.) or greater in
thickness shall be determined according to Test Method D 638.
1.2 This test method may be used to test all plastics within
the thickness range described and the capacity of the machine
employed.
1.2.1 Static Weighing, Constant-Rate-of-Grip Separation
Test—This test method employs a constant rate of separation of
the grips holding the ends of the test specimen.
1.3 Specimen extension may be measured in these test
methods by grip separation, extension indicators, or displace-
ment of gage marks.
1.4 A procedure for determining the tensile modulus of
elasticity is included at one strain rate.
NOTE 3—The modulus determination is generally based on the use of
grip separation as a measure of extension; however, the desirability of
using extensometers, as described in 5.2, is recognized and provision for
the use of such instrumentation is incorporated in the procedure.
1.5 Test data obtained by this test method is relevant and
appropriate for use in engineering design.
1.6 The values stated in SI units are to be regarded as the
standard. The values in parentheses are provided for informa-
tion only.
1.7 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
NOTE 4—This test method is similar to ISO 527-3, but is not considered
technically equivalent. ISO 527-3 allows for additional specimen configu-
rations, specifies different test speeds, and requires an extensometer or
gage marks on the specimen.
2. Referenced Documents
2.1 ASTM Standards:2
D 618 Practice for Conditioning Plastics for Testing
D 638 Test Method for Tensile Properties of Plastics
D 4000 Classification System for Specifying Plastic Mate-
rials
D 5947 Test Methods for Physical Dimensions of Solid
Plastics Specimens
D 6287 Practice for Cutting Film and Sheeting Test Speci-
mens
E 4 Practices for Force Verification of Testing Machines
E 691 Practice for Conducting an Interlaboratory Study to
Determine the Precision of a Test Method
2.2 ISO Standard:
ISO 527-3 Plastics—Determination of Tensile Properties—
Part 3: Test Conditions for Films and Sheets3
3. Terminology
3.1 Definitions—Definitions of terms and symbols relating
to tension testing of plastics appear in the Annex to Test
Method D 638.
3.1.1 line grips—grips having faces designed to concentrate
the entire gripping force along a single line perpendicular to the
direction of testing stress. This is usually done by combining
one standard flat face and an opposing face from which
protrudes a half-round.
3.1.2 tear failure—a tensile failure characterized by fracture
initiating at one edge of the specimen and progressing across
the specimen at a rate slow enough to produce an anomalous
load-deformation curve.
1 These test methods are under the jurisdiction of ASTM Committee D20 on
Plastics and are the direct responsibility of Subcommittee D20.19 on Film and
Sheeting.
Current edition approved Jan. 1, 2009. Published January 2009. Originally
approved in 1946. Last previous edition approved in 2002 as D 882 - 02.
2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
3 Available from American National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036, http://www.ansi.org.
1
*A Summary of Changes section appears at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
4. Significance and Use
4.1 Tensile properties determined by this test method are of
value for the identification and characterization of materials for
control and specification purposes. Tensile properties may vary
with specimen thickness, method of preparation, speed of
testing, type of grips used, and manner of measuring extension.
Consequently, where precise comparative results are desired,
these factors must be carefully controlled. This test method
shall be used for referee purposes, unless otherwise indicated
in particular material specifications. For many materials, there
may be a specification that requires the use of this test method,
but with some procedural modifications that take precedence
when adhering to the specification. Therefore, it is advisable to
refer to that material specification before using this test
method. Table 1 in Classification D 4000 lists the ASTM
materials standards that currently exist.
4.2 Tensile properties may be utilized to provide data for
research and development and engineering design as well as
quality control and specification. However, data from such
tests cannot be considered significant for applications differing
widely from the load-time scale of the test employed.
4.3 The tensile modulus of elasticity is an index of the
stiffness of thin plastic sheeting. The reproducibility of test
results is good when precise control is maintained over all test
conditions. When different materials are being compared for
stiffness, specimens of identical dimensions must be employed.
4.4 The tensile energy to break (TEB) is the total energy
absorbed per unit volume of the specimen up to the point of
rupture. In some texts this property has been referred to as
toughness. It is used to evaluate materials that may be
subjected to heavy abuse or that might stall web transport
equipment in the event of a machine malfunction in end-use
applications. However, the rate of strain, specimen parameters,
and especially flaws may cause large variations in the results.
In that sense, caution is advised in utilizing TEB test results for
end-use design applications.
4.5 Materials that fail by tearing give anomalous data which
cannot be compared with those from normal failure.
5. Apparatus
5.1 Testing Machine—A testing machine of the constant
rate-of-crosshead-movement type and comprising essentially
the following:
5.1.1 Fixed Member—A fixed or essentially stationary
member carrying one grip.
5.1.2 Movable Member—A movable member carrying a
second grip.
5.1.3 Grips—A set of grips for holding the test specimen
between the fixed member and the movable member of the
testing machine; grips can be either the fixed or self-aligning
type. In either case, the gripping system must minimize both
slippage and uneven stress distribution.
5.1.3.1 Fixed grips are rigidly attached to the fixed and
movable members of the testing machine. When this type of
grip is used, care must be taken to ensure that the test specimen
is inserted and clamped so that the long axis of the test
specimen coincides with the direction of pull through the
center line of the grip assembly.
5.1.3.2 Self-aligning grips are attached to the fixed and
movable members of the testing machine in such a manner that
they will move freely into alignment as soon as a load is
applied so that the long axis of the test specimen will coincide
with the direction of the applied pull through the center line of
the grip assembly. The specimens should be aligned as per-
fectly as possible with the direction of pull so that no rotary
motion that may induce slippage will occur in the grips; there
is a limit to the amount of misalignment self-aligning grips will
accommodate.
5.1.3.3 The test specimen shall be held in such a way that
slippage relative to the grips is prevented insofar as possible.
Grips lined with thin rubber, crocus-cloth, or pressure-sensitive
tape as well as file-faced or serrated grips have been success-
fully used for many materials. The choice of grip surface will
depend on the material tested, thickness, etc. Line grips padded
on the round face with 1.0 mm (40 mil) blotting paper or filter
paper have been found superior. Air-actuated grips have been
found advantageous, particularly in the case of materials that
tend to “neck” into the grips, since pressure is maintained at all
times. In cases where samples frequently fail at the edge of the
grips, it may be advantageous to increase slightly the radius of
curvature of the edges where the grips come in contact with the
test area of the specimen.
5.1.4 Drive Mechanism—A drive mechanism for imparting
to the movable member a uniform, controlled velocity with
respect to the stationary member. The velocity shall be regu-
lated as specified in Section 9.
5.1.5 Load Indicator—A suitable load-indicating mecha-
nism capable of showing the total tensile load carried by the
test specimen held by the grips. This mechanism shall be
essentially free of inertial lag at the specified rate of testing (see
Note 5). Unless a suitable extensometer is used (see 5.2), the
motion of the weighing system shall not exceed 2 % of the
specimen extension within the range being measured. The load
indicator shall determine the tensile load applied to the
specimen with an accuracy of 61 % of the indicated value, or
better. The accuracy of the testing machine shall be verified in
accordance with Practices E 4.
5.1.6 Crosshead Extension Indicator—A suitable extension-
indicating mechanism capable of showing the amount of
change in the separation of the grips, that is, crosshead
movement. This mechanism shall be essentially free of inertial
lag at the specified rate of testing (see Note 5) and shall
indicate the crosshead movement with an accuracy of 61 % of
the indicated value, or better.
5.2 Extensometer (Optional)—A suitable instrument may, if
desired, be used for determining the distance between two
designated points on the test specimen as the specimen is
stretched. This apparatus, if employed, shall be so designed as
to minimize stress on the specimen at the contact points of the
specimen and the instrument (see 8.3). It is desirable that this
instrument automatically record the distance, or any change in
it, as a function of the load on the test specimen or of the
elapsed time from the start of the test, or both. If only the latter
is obtained, load-time data must also be taken. This instrument
must be essentially free of inertial lag at the specified speed of
testing (see Note 5).
D 882 – 09
2
5.2.1 Modulus of Elasticity and Low-Extension
Measurements—Extensometers used for modulus of elasticity
and low-extension (less than 20 % elongation) measurements
shall, at a minimum, be accurate to 61 % and comply with the
requirements set forth in Practice E 83 for a Class C instru-
ment.
5.2.2 High-Extension Measurements—Instrumentation and
measuring techniques used for high-extension (20 % elonga-
tion or greater) measurements shall be accurate to 610 % of
the indicated value, or better.
NOTE 5—A sufficiently high response speed in the indicating and
recording system for the load and extension data is essential. The response
speed required of the system will depend in part on the material tested
(high or low elongation) and the rate of straining.
5.3 Thickness Gage—A dead-weight dial micrometer as
prescribed in Method C of Test Methods D 5947, or an
equivalent measuring device, reading to 0.0025 mm (0.0001
in.) or less.
5.4 Width-Measuring Devices—Suitable test scales or other
width measuring devices capable of measuring 0.25 mm (0.010
in.) or less.
5.5 Specimen Cutter—For the apparatus and techniques for
cutting film and sheeting used in this test method, refer to
Practice D 6287.
5.5.1 Devices that use razor blades have proven especially
suitable for materials having an elongation-at-fracture above
10 to 20 %.
5.5.2 The use of punch press or striking dies are not
recommended because poor and inconsistent specimen edges
may be produced.
6. Test Specimens
6.1 The test specimens shall consist of strips of uniform
width and thickness at least 50 mm (2 in.) longer than the grip
separation used.
6.2 The nominal width of the specimens shall be not less
than 5.0 mm (0.20 in.) or greater than 25.4 mm (1.0 in.).
6.3 A width-thickness ratio of at least eight shall be used.
Narrow specimens magnify effects of edge strains or flaws, or
both.
6.4 The utmost care shall be exercised in cutting specimens
to prevent nicks and tears which are likely to cause premature
failures (Note 6). The edges shall be parallel to within 5 % of
the width over the length of the specimen between the grips.
NOTE 6—Microscopical examination of specimens may be used to
detect flaws due to sample or specimen preparation.
6.5 Wherever possible, the test specimens shall be selected
so that thickness is uniform to within 10 % of the thickness
over the length of the specimen between the grips in the case
of materials 0.25 mm (0.010 in.) or less in thickness and to
within 5 % in the case of materials greater than 0.25 mm (0.010
in.) in thickness but less than 1.00 mm (0.040 in.) in thickness.
NOTE 7—In cases where thickness variations are in excess of those
recommended in 6.5, results may not be characteristic of the material
under test.
6.6 If the material is suspected of being anisotropic, two sets
of test specimens shall be prepared having their long axes
respectively parallel with and normal to the suspected direction
of anisotropy.
6.7 For tensile modulus of elasticity determinations, a
specimen gage length of 250 mm (10 in.) shall be considered
as standard. This length is used in order to minimize the effects
of grip slippage on test results. When this length is not feasible,
test sections as short as 100 mm (4 in.) may be used if it has
been shown that results are not appreciably affected. However,
the 250-mm gage length shall be used for referee purposes. The
speed of testing of shorter specimens must be adjusted in order
for the strain rate to be equivalent to that of the standard
specimen.
NOTE 8—Two round robin tests4 have shown that, for materials of less
than 0.25-mm (10-mil) thickness, line grips padded on the round side with
1.0-mm (40-mil) blotting paper give the same results with a 100-mm test
section as a 250-mm test section produces with flat-face grips.
NOTE 9—Excessive jaw slippage becomes increasingly difficult to
overcome in cases where high modulus materials are tested in thicknesses
greater than 0.25 mm (0.010 in.).
7. Conditioning
7.1 Conditioning—Condition the test specimens at 23 6
2°C (73.4 6 3.6°F) and 50 6 10 % relative humidity for not
less than 40 h prior to test in accordance with Procedure A of
Practice D 618 unless otherwise specified by agreement or the
relevant ASTM material specification. In cases of disagree-
ment, the tolerances shall be 61°C (61.8°F) and 65 %
relative humidity.
7.2 Test Conditions—Conduct the tests at 23 6 2°C (73.4 6
3.6°F) and 50 6 10 % relative humidity unless otherwise
specified by agreement or the relevant ASTM material speci-
fication. In cases of disagreement, the tolerances shall be 61°C
(61.8°F) and 65 % relative humidity.
8. Number of Test Specimens
8.1 In the case of isotropic materials, at least five specimens
shall be tested from each sample.
8.2 In the case of anisotropic materials, at least ten speci-
mens, five normal and five parallel with the principal axis of
anisotropy, shall be tested from each sample.
8.3 Specimens that fail at some obvious flaw or that fail
outside the gage length shall be discarded and retests made,
unless such flaws or conditions constitute a variable whose
effect is being studied. However, jaw breaks (failures at the
grip contact point) are acceptable if it has been shown that
results from such tests are in essential agreement with values
obtained from breaks occurring within the gage length.
NOTE 10—In the case of some materials, examination of specimens,
prior to and following testing, under crossed optical polarizers (polarizing
films) provides a useful means of detecting flaws which may be, or are,
responsible for premature failure.
4 Supporting data are available from ASTM Headquarters. Request RR: D20-
1058.
D 882 – 09
3
9. Speed of Testing
9.1 The speed of testing is the rate of separation of the two
members (or grips) of the testing machine when running idle
(under no load). This rate of separation shall be maintained
within 5 % of the no-load value when running under full-
capacity load.
9.2 The speed of testing shall be calculated from the
required initial strain rate as specified in Table 1. The rate of
grip separation may be determined for the purpose of these test
methods from the initial strain rate as follows:
A 5 BC (1)
where:
A = rate of grip separation, mm (or in.)/min,
B = initial distance between grips, mm (or in.), and
C = initial strain rate, mm/mm·min (or in./in.·min).
9.3 The initial strain rate shall be as in Table 1 unless
otherwise indicated by the specification for the material being
tested.
NOTE 11—Results obtained at different initial strain rates are not
comparable; consequently, where direct comparisons between materials in
various elongation classes are required, a single initial strain rate should
be used. For some materials it may be advisable to select the strain rates
on the basis of percent elongation at yield.
9.4 In cases where conflicting material classification, as
determined by percent elongation at break values, results in a
choice of strain rates, the lower rate shall be used.
9.5 If modulus values are being determined, separate speci-
mens shall be used whenever strain rates and specimen
dimensions are not the same as those employed in the test for
other tensile properties.
10. Procedure
10.1 Select a load range such that specimen failure occurs
within its upper two thirds. A few trial runs may be necessary
to select a proper combination of load range and specimen
width.
10.2 Measure the cross-sectional area of the specimen at
several points along its length. Measure the width to an
accuracy of 0.25 mm (0.010 in.) or better. Measure the
thickness to an accuracy of 0.0025 mm (0.0001 in.) or better
for films less than 0.25 mm (0.010 in.) in thickness and to an
accuracy of 1 % or better for films greater than 0.25 mm (0.010
in.) but less than 1.0 mm (0.040 in.) in thickness.
10.3 Set the initial grip separation in accordance with Table
1.
10.4 Set the rate of grip separation to give the desired strain
rate, based on the initial distance between the grips, in
accordance with Table 1. Zero the calibrated load weighing
system, extension indicator(s) and recording system.
NOTE 12—Extensometers may be used for modulus of elasticity deter-
minations with the expectation of obtaining more accurate values than
may be obtained using grip separation as the effective gage length.
Precautions should be taken to ensure that extensometer slippage and
undue stressing of the specimen do not occur. Refer also to 6.7.
10.5 In cases where it is desired to measure a test section
other than the total length between the grips, mark the ends of
the desired test section with a soft, fi
本文档为【ASTM D882-09B】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。