Designation: D 445 – 06
Designation: 71/1/97
An American National Standard
British Standard 2000: Part 71:1990
Standard Test Method for
Kinematic Viscosity of Transparent and Opaque Liquids
(and Calculation of Dynamic Viscosity)1
This standard is issued under the fixed designation D 445; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
This standard has been approved for use by agencies of the Department of Defense.
1. Scope*
1.1 This test method specifies a procedure for the determi-
nation of the kinematic viscosity, n, of liquid petroleum
products, both transparent and opaque, by measuring the time
for a volume of liquid to flow under gravity through a
calibrated glass capillary viscometer. The dynamic viscosity, h,
can be obtained by multiplying the kinematic viscosity, n, by
the density, r, of the liquid.
NOTE 1—For the measurement of the kinematic viscosity and viscosity
of bitumens, see also Test Methods D 2170 and D 2171.
NOTE 2—ISO 3104 corresponds to Test Method D 445.
1.2 The result obtained from this test method is dependent
upon the behavior of the sample and is intended for application
to liquids for which primarily the shear stress and shear rates
are proportional (Newtonian flow behavior). If, however, the
viscosity varies significantly with the rate of shear, different
results may be obtained from viscometers of different capillary
diameters. The procedure and precision values for residual fuel
oils, which under some conditions exhibit non-Newtonian
behavior, have been included.
1.3 The range of kinematic viscosities covered by this test
method is from 0.2 to 300 000 mm2/s (see Table A1.1) at all
temperatures (see 6.3 and 6.4). The precision has only been
determined for those materials, kinematic viscosity ranges and
temperatures as shown in the footnotes to the precision section.
1.4 The values stated in SI units are to be regarded as the
standard. The values given in parentheses are for information
only.
1.5 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards: 2
D 446 Specifications and Operating Instructions for Glass
Capillary Kinematic Viscometers
D 1193 Specification for Reagent Water
D 1217 Test Method for Density and Relative Density
(Specific Gravity) of Liquids by Bingham Pycnometer
D 1480 Test Method for Density and Relative Density
(Specific Gravity) of Viscous Materials by Bingham Pyc-
nometer
D 1481 Test Method for Density and Relative Density
(Specific Gravity) of Viscous Materials by Lipkin Bicap-
illary Pycnometer
D 2162 Practice for Basic Calibration of Master Viscom-
eters and Viscosity Oil Standards
D 2170 Test Method for Kinematic Viscosity of Asphalts
(Bitumens)
D 2171 Test Method for Viscosity of Asphalts by Vacuum
Capillary Viscometer
D 6071 Test Method for Low Level Sodium in High Purity
Water by Graphite Furnace Atomic Absorption Spectros-
copy
D 6074 Guide for Characterizing Hydrocarbon Lubricant
Base Oils
D 6617 Practice for Laboratory Bias Detection Using
Single Test Result from Standard Material
E 1 Specification for ASTM Liquid-in-Glass Thermometers
E 77 Test Method for Inspection and Verification of Ther-
mometers
2.2 ISO Standards:3
1 This test method is under the jurisdiction of ASTM Committee D02 on
Petroleum Products and Lubricants and is the direct responsibility of Subcommittee
D02.07 on Flow Properties.
Current edition approved May 15, 2006. Published June 2006. Originally
approved in 1937. Last previous edition approved in 2004 as D 445–04e2.
In the IP, this test method is under the jurisdiction of the Standardization
Committee.
2 For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
3 Available from American National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036.
1
*A Summary of Changes section appears at the end of this standard.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Copyright ASTM International
Provided by IHS under license with ASTM Licensee=North Carolina DOT/5959838001
Not for Resale, 07/22/2007 01:10:05 MDTNo reproduction or networking permitted without license from IHS
-
-
`
`
`
,
`
`
,
`
,
,
`
`
,
`
`
`
`
`
`
`
,
`
`
`
,
`
`
`
-
`
-
`
,
,
`
,
,
`
,
`
,
,
`
-
-
-
ISO 3104 Petroleum Products—Transparent and Opaque
Liquids—Determination of Kinematic Viscosity and Cal-
culation of Dynamic Viscosity
ISO 3105 Glass Capillary Kinematic Viscometers—
Specification and Operating Instructions
ISO 3696 Water for Analytical Laboratory Use—
Specification and Test Methods
ISO 5725 Accuracy (trueness and precision) of measure-
ment methods and results.
ISO 9000 Quality Management and Quality Assurance
Standards—Guidelines for Selection and Use
ISO 17025 General Requirements for the Competence of
Testing and Calibration Laboratories
2.3 NIST Standards:4
NIST Technical Note 1297, Guideline for Evaluating and
Expressing the Uncertainty of NIST Measurement Results
NIST GMP 11
NIST Special Publication 819
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 automated viscometer, n—apparatus which, in part or
in whole, has mechanized one or more of the procedural steps
indicated in Section 11 or 12 without changing the principle or
technique of the basic manual apparatus. The essential ele-
ments of the apparatus in respect to dimensions, design, and
operational characteristics are the same as those of the manual
method.
3.1.1.1 Discussion—Automated viscometers have the capa-
bility to mimic some operation of the test method while
reducing or removing the need for manual intervention or
interpretation. Apparatus which determine kinematic viscosity
by physical techniques that are different than those used in this
test method are not considered to be Automated Viscometers.
3.1.2 density, n—the mass per unit volume of a substance at
a given temperature.
3.1.3 dynamic viscosity, n—the ratio between the applied
shear stress and rate of shear of a liquid.
3.1.3.1 Discussion—It is sometimes called the coefficient of
dynamic viscosity or, simply, viscosity. Thus dynamic viscos-
ity is a measure of the resistance to flow or deformation of a
liquid.
3.1.3.2 Discussion—The term dynamic viscosity can also
be used in a different context to denote a frequency-dependent
quantity in which shear stress and shear rate have a sinusodial
time dependence.
3.1.4 kinematic viscosity, n—the resistance to flow of a fluid
under gravity.
3.1.4.1 Discussion—For gravity flow under a given hydro-
static head, the pressure head of a liquid is proportional to its
density, r. For any particular viscometer, the time of flow of a
fixed volume of fluid is directly proportional to its kinematic
viscosity, n, where n = h/r, and h is the dynamic viscosity
coefficient.
4. Summary of Test Method
4.1 The time is measured for a fixed volume of liquid to
flow under gravity through the capillary of a calibrated
viscometer under a reproducible driving head and at a closely
controlled and known temperature. The kinematic viscosity
(determined value) is the product of the measured flow time
and the calibration constant of the viscometer. Two such
determinations are needed from which to calculate a kinematic
viscosity result that is the average of two acceptable deter-
mined values.
5. Significance and Use
5.1 Many petroleum products, and some non-petroleum
materials, are used as lubricants, and the correct operation of
the equipment depends upon the appropriate viscosity of the
liquid being used. In addition, the viscosity of many petroleum
fuels is important for the estimation of optimum storage,
handling, and operational conditions. Thus, the accurate deter-
mination of viscosity is essential to many product specifica-
tions.
6. Apparatus
6.1 Viscometers—Use only calibrated viscometers of the
glass capillary type, capable of being used to determine
kinematic viscosity within the limits of the precision given in
the precision section.
6.1.1 Viscometers listed in Table A1.1, whose specifications
meet those given in Specifications D 446 and in ISO 3105 meet
these requirements. It is not intended to restrict this test method
to the use of only those viscometers listed in Table A1.1. Annex
A1 gives further guidance.
6.1.2 Automated Viscometers—Automated apparatus may
be used as long as they mimic the physical conditions,
operations or processes of the manual apparatus. Any viscom-
eter, temperature measuring device, temperature control, tem-
perature controlled bath or timing device incorporated in the
automated apparatus shall conform to the specification for
these components as stated in Section 6 of this test method.
Flow times of less than 200 s are permitted, however, a kinetic
energy correction shall be applied in accordance with Section
7 on Kinematic Viscosity Calculation of Specifications D 446.
The kinetic energy correction shall not exceed 3.0 % of the
measured viscosity. The automated apparatus shall be capable
of determining kinematic viscosity of a certified viscosity
reference standard within the limits stated in 9.2.1 and Section
17. The precision shall be of statistical equivalence to, or better
(has less variability) than the manual apparatus.
NOTE 3—Precision and bias of kinematic viscosity measurements for
flow times of less than 200 s has not been determined. The precision stated
in Section 17 is not know to be valid for kinematic viscosity measure-
ments with flow times less than 200 s.
6.2 Viscometer Holders—Use viscometer holders to enable
all viscometers which have the upper meniscus directly above
the lower meniscus to be suspended vertically within 1° in all
directions. Those viscometers whose upper meniscus is offset
from directly above the lower meniscus shall be suspended
vertically within 0.3° in all directions (see Specifications D 446
and ISO 3105).
4 Available from National Institute of Standards and Technology (NIST), 100
Bureau Dr., Stop 3460, Gaithersburg, MD 20899-3460.
D 445 – 06
2Copyright ASTM International
Provided by IHS under license with ASTM Licensee=North Carolina DOT/5959838001
Not for Resale, 07/22/2007 01:10:05 MDTNo reproduction or networking permitted without license from IHS
--```,``,`,,``,```````,```,```-`-`,,`,,`,`,,`---
6.2.1 Viscometers shall be mounted in the constant tempera-
ture bath in the same manner as when calibrated and stated on
the certificate of calibration. See Specifications D 446, see
Operating Instructions in Annexes A1–A3. For those viscom-
eters which have Tube L (see Specifications D 446) held
vertical, vertical alignment shall be confirmed by using (1) a
holder ensured to hold Tube L vertical, or (2) a bubble level
mounted on a rod designed to fit into Tube L, or (3) a plumb
line suspended from the center of Tube L, or (4) other internal
means of support provided in the constant temperature bath.
6.3 Temperature-Controlled Bath—Use a transparent liquid
bath of sufficient depth such, that at no time during the
measurement of flow time, any portion of the sample in the
viscometer is less than 20 mm below the surface of the bath
liquid or less than 20 mm above the bottom of the bath.
6.3.1 Temperature Control—For each series of flow time
measurements, the temperature control of the bath liquid shall
be such that within the range from 15 to 100°C, the tempera-
ture of the bath medium does not vary by more than 60.02°C
of the selected temperature over the length of the viscometer,
or between the position of each viscometer, or at the location of
the thermometer. For temperatures outside this range, the
deviation from the desired temperature must not exceed
60.05°C.
6.4 Temperature Measuring Device in the Range from 0 to
100°C—Use either calibrated liquid-in-glass thermometers
(Annex A2) of an accuracy after correction of 60.02°C or
better, or any other thermometric device of equal or better
accuracy.
6.4.1 If calibrated liquid-in-glass thermometers are used, the
use of two thermometers is recommended. The two thermom-
eters shall agree within 0.04°C.
6.4.2 Outside the range from 0 to 100°C, use either cali-
brated liquid-in-glass thermometers of an accuracy after cor-
rection of 60.05°C or better, or any other thermometric device
of equal or better accuracy. When two temperature measuring
devices are used in the same bath, they shall agree within
60.1°C.
6.4.3 When using liquid-in-glass thermometers, such as
those in Table A2.1, use a magnifying device to read the
thermometer to the nearest 1⁄5 division (for example, 0.01°C or
0.02°F) to ensure that the required test temperature and
temperature control capabilities are met (see 10.1). It is
recommended that thermometer readings (and any corrections
supplied on the certificates of calibrations for the thermom-
eters) be recorded on a periodic basis to demonstrate compli-
ance with the test method requirements. This information can
be quite useful, especially when investigating issues or causes
relating to testing accuracy and precision.
6.5 Timing Device—Use any timing device that is capable
of taking readings with a discrimination of 0.1 s or better and
has an accuracy within 60.07 % (see Annex A3) of the reading
when tested over the minimum and maximum intervals of
expected flow times.
6.5.1 Electrical timing devices may be used if the current
frequency is controlled to an accuracy of 0.05 % or better.
Alternating currents, as provided by some public power sys-
tems, are intermittently rather than continuously controlled.
When used to actuate electrical timing devices, such control
can cause large errors in kinematic viscosity flow time mea-
surements.
7. Reagents and Materials
7.1 Chromic Acid Cleaning Solution, or a nonchromium-
containing, strongly oxidizing acid cleaning solution.
(Warning—Chromic acid is a health hazard. It is toxic, a
recognized carcinogen, highly corrosive, and potentially haz-
ardous in contact with organic materials. If used, wear a full
face-shield and full-length protective clothing including suit-
able gloves. Avoid breathing vapor. Dispose of used chromic
acid carefully as it remains hazardous. Nonchromium-
containing, strongly oxidizing acid cleaning solutions are also
highly corrosive and potentially hazardous in contact with
organic materials, but do not contain chromium which has
special disposal problems.)
7.2 Sample Solvent, completely miscible with the sample.
Filter before use.
7.2.1 For most samples a volatile petroleum spirit or naph-
tha is suitable. For residual fuels, a prewash with an aromatic
solvent such as toluene or xylene may be necessary to remove
asphaltenic material.
7.3 Drying Solvent, a volatile solvent miscible with the
sample solvent (see 7.2) and water (see 7.4). Filter before use.
7.3.1 Acetone is suitable. (Warning—Extremely flam-
mable.)
7.4 Water, deionized or distilled and conforming to Speci-
fication D 1193 or Grade 3 of ISO 3696. Filter before use.
8. Certified Viscosity Reference Standards
8.1 Certified viscosity reference standards shall be certified
by a laboratory that has been shown to meet the requirements
of ISO 17025 by independent assessment. Viscosity standards
shall be traceable to master viscometer procedures described in
Test Method D 2162.
8.2 The uncertainty of the certified viscosity reference
standard shall be stated for each certified value (k = 2, 95%
confidence). See ISO 5725 or NIST 1297.
9. Calibration and Verification
9.1 Viscometers—Use only calibrated viscometers, ther-
mometers, and timers as described in Section 6.
9.2 Certified Viscosity Reference Standards (Table A1.2)—
These are for use as confirmatory checks on the procedure in
the laboratory.
9.2.1 If the determined kinematic viscosity does not agree
within the acceptable tolerance band, as calculated from Annex
A4, of the certified value, recheck each step in the procedure,
including thermometer and viscometer calibration, to locate the
source of error. Annex A1 gives details of standards available.
NOTE 4—In previous issues of Test Method D 445, limits of 60.35% of
the certified value have been used. The data to support the limit of
60.35% cannot be verified. Annex A4 provides instructions on how to
determine the tolerance band. The tolerance band combines both the
uncertainty of the certified viscosity reference standard as well as the
uncertainty of the laboratory using the certified viscosity reference
standard.
D 445 – 06
3Copyright ASTM International
Provided by IHS under license with ASTM Licensee=North Carolina DOT/5959838001
Not for Resale, 07/22/2007 01:10:05 MDTNo reproduction or networking permitted without license from IHS
-
-
`
`
`
,
`
`
,
`
,
,
`
`
,
`
`
`
`
`
`
`
,
`
`
`
,
`
`
`
-
`
-
`
,
,
`
,
,
`
,
`
,
,
`
-
-
-
9.2.1.1 As an alternative to the calculation in Annex A4, the
approximate tolerance bands in Table 1 may be used.
9.2.2 The most common sources of error are caused by
particles of dust lodged in the capillary bore and temperature
measurement errors. It must be appreciated that a correct result
obtained on a standard oil does not preclude the possibility of
a counterbalancing combination of the possible sources of
error.
9.3 The calibration constant, C, is dependent upon the
gravitational acceleration at the place of calibration and this
must, therefore, be supplied by the standardization laboratory
together with the instrument constant. Where the acceleration
of gravity, g, differs by more that 0.1 %, correct the calibration
constant as follows:
C2 5 ~g2/g1! 3 C1 (1)
where the subscripts 1 and 2 indicate, respectively, the
standardization laboratory and the testing laboratory.
10. General Procedure for Kinematic Viscosity
10.1 Adjust and maintain the viscometer bath at the required
test temperature within the limits given in 6.3.1 taking account
of the conditions given in Annex A2 and of the corrections
supplied on the certificates of calibration for the thermometers.
10.1.1 Thermometers shall be held in an upright position
under the same conditions of immersion as when calibrated.
10.1.2 In order to obtain the most reliable temperature
measurement, it is recommended that two thermometers with
valid calibration certificates be used (see 6.4).
10.1.3 They should be viewed with a lens assembly giving
approximately five times magnification and be arranged to
eliminate parallax errors.
10.2 Select a clean, dry, calibrated viscometer having a
range covering the estimated kinematic viscosity (that is, a
wide capillary for a very viscous liquid and a narrower
capillary for a more fluid liquid). The flow time for manual
viscometers shall not be less than 200 s or the longer time
noted in Specifications D 446. Flow times of less than 200 s are
permitted for automated viscometers, provided they meet the
requirements of 6.1.2.
10.2.1 The specific details of operation vary for the different
types of viscometers listed in Table A1.1. The operating
instructions for the different types of viscometers are given in
Specifications D 446.
10.2.2 When the test temperature is below the dew point, fill
the viscometer in the normal manner as required in 11.1. To
ensure that moisture does not condense or freeze on the walls
of the capillary, draw the test portion into the working capillary
and timi
本文档为【ASTM D 445-06】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。