[ Care and Use ManUal ]
Waters Spherisorb Columns 1
Waters spherisorb columns
contents
i. GettinG started
a. Column Installation
1. Reversed-Phase Columns
2. Normal-Phase Columns
b. Column Equilibration
1. Reversed-Phase Columns
2. Normal-Phase Columns
c. Initial Column Efficiency Determination
ii. column use
a. Guard Columns
b. Sample Preparation
c. pH Range
d. Solvents
e. Pressure
f. Temperature
iii. scalinG up/doWn isocratic methods
iv. troubleshootinG
v. column cleaninG, reGeneratinG and storaGe
a. Cleaning and Regenerating
1. Reversed-Phase Columns
2. Normal-Phase Columns
b. Storage
1. Reversed-Phase Columns
2. Normal-Phase Columns
vi. connectinG the column to the hplc
a. Column Connectors and System Tubing Considerations
b. Band Spreading Minimization
c. Measuring System Bandspreading Volume & System Variance
d. Measuring System Volume
vii. additional inFormation
a. Use of Narrow-Bore (≤3.0 mm i.d.) Columns
b. Impact of Bandspreading Volume on 2.1 mm i.d. Column Performance
c. Non-Optimized vs. Optimized LC/MS/MS System:
System Modification Recommendations
d. Guard Cartridges and Columns Assembly
Thank you for choosing a Waters Spherisorb® column. Spherisorb
analytical columns are durable, high-efficiency chromatographic
columns, featured in thousands of references in chromatographic
literature. The wide range of column lengths and diameters offers
you exceptional flexibility in optimizing methods and reducing
solvent consumption. Follow the guidelines in this manual to obtain
the best performance, reproducibility and durability from your
analytical columns and cartridges.
[ Care and Use ManUal ]
Waters Spherisorb Columns 2
Chemistry
Particle
Shape
Particle Size
(µm)
Pore Size
(Å)
Surface Area
(m2/g)
Pore Volume
(cc/g)
% Carbon
Load
Endcapped
Silica Spherical 3, 5 and 10 80 220 0.50 n/a n/a
ODS2 (C18) - Fully End Capped Spherical 3, 5 and 10 80 220 0.50 11.5 yes
ODS1 (C18) - Partially End Capped Spherical 3, 5 and 10 80 220 0.50 6.2 no
ODSB (C18) - Base De-activated Spherical 5 80 220 0.50 11.5 yes*
C8 Spherical 3, 5 and 10 80 220 0.50 5.8 yes
C6 Spherical 3, 5 and 10 80 220 0.50 4.7 yes
C1 Spherical 3, 5 and 10 80 220 0.50 2.2 no
Nitrile (CN) Spherical 3, 5 and 10 80 220 0.50 3.1 no
Amino (NH2) Spherical 3, 5 and 10 80 220 0.50 1.9 no
Phenyl Spherical 3, 5 and 10 80 220 0.50 2.5 no
OD/CN (Mixed Mode) Spherical 5 80 220 0.50 5.0 yes
SAX Spherical 3, 5 and 10 80 220 0.50 4.0 no
SCX Spherical 3, 5 and 10 80 220 0.50 4.0 no
i . Get t inG start ed
Each Spherisorb column comes with a Performance Test
Chromatogram. This Performance Test Chromatogram is specific
to each individual column and contains the following information:
gel batch number, column serial number, USP plate count, USP
tailing factor, capacity factor, and chromatographic conditions.
The performance test chromatogram should be stored for future
reference.
a. Column Installation
Note: The flow rates given in the procedure below are for a typical
4.6 mm i.d. column. Scale the flow rate up or down accordingly based
upon the column i.d., length, particle size, and backpressure of the
Spherisorb column being installed. See “Scaling Up/Down Isocratic
Separations” for calculating flow rates when changing column i.d.
and/or length. See “Connecting the Column to the HPLC” for a more
detailed discussion on HPLC connections.
1. Reversed-Phase Columns
1. Purge the pumping system of any buffer-containing mobile phases and
connect the inlet end of the column to the injector outlet.
2. Flush column with 100% organic mobile phase (methanol or
acetonitrile) by setting the pump flow rate to 0.1 mL/min and increase
the flow rate to 1 mL/min over 5 minutes.
3. When the mobile phase is flowing freely from the column outlet, stop
the flow and attach the column outlet to the detector. This prevents
entry of air into the detection system and gives more rapid baseline
equilibration.
4. Gradually increase the flow rate as described in step 2.
5. Once a steady backpressure and baseline have been achieved, proceed
to the next section.
2. Normal-Phase Columns
Note: It is assumed that your system has been used for reversed-phase
chromatography. If this is not the case, you can start with step 3.
1. Purge the pumping system of any buffer containing mobile phases.
2. Flush the system thoroughly with acetonitrile.
3. Switch the system over to the mobile phase that you are planning to use
in normal-phase chromatography.
4. Connect the column and equilibrate it with the mobile phase.
Note: Equilibration with the mobile phase may require a larger amount of
solvent than in reversed-phase chromatography.
Table 1. Spherisorb Column Physical Characteristics
* polar endcapping
[ Care and Use ManUal ]
Waters Spherisorb Columns 3
b. Column Equilibration
Spherisorb columns are shipped in test mobile phase. It is important
to ensure mobile phase compatibility before changing to a different
mobile phase system. Equilibrate the column with a minimum of 10
column volumes of the mobile phase to be used (refer to Table 2 for
a listing of empty column volumes).
1. Reversed-Phase Columns
To avoid precipitating out mobile-phase buffers on your column
or in your system, flush the column with five column volumes of
a water/organic solvent mixture, using the same or lower solvent
content as in the desired buffered mobile phase. (For example, flush
the column and HPLC system with 60% methanol in water prior to
introducing 60% methanol/40% buffer mobile phase.)
Note: If mobile phase additives are present in low concentrations (e.g., ion-pairing
reagents), 100 to 200 column volumes may be required for complete equilibration.
2. Normal-Phase (Spherisorb Silica, Amino, Cyano) Column
Spherisorb normal-phase (NP) columns are delivered in 96% heptane /
4% isopropyl alcohol. Care should be taken not to pass any mobile
phase through the column that might cause a precipitate. Spherisorb
NP columns are compatible with water and all common organic
solvents, provided that solvent miscibility is accounted for.
Equilibrate normal-phase silica columns in the mobile phase. Very
small quantities of water in the mobile phase can dramatically affect
the activity of normal-phase packings. For good reproducibility,
ensure that the mobile phase always has the same water content.
It is difficult and usually unnecessary to completely eliminate the
water from the mobile phase. Dry mobile phases can take a very long
time to equilibrate the column. A water content of 50 percent of
saturation is recommended for most applications.
To equilibrate your column:
1. Starting at 0.0 mL/min, increase the flow rate in 0.1 mL/min
increments to 1.0 minutes.
2. Purge the column with the mobile phase until you obtain a stable
baseline.
3. Verify that retention times and peak areas for a standard are stable
by comparing 2-3 replicate consecutive injections
Before you perform the first analysis on your new column, perform
an efficiency test to confirm the performance of the column.
Table 2. Empty Column Volumes in mL
(multiply by 10 for flush solvent volumes)
Column
Length
Column Internal Diameter (mm)
1.0 2.1 3.0 4.6 10 20
20 mm - 0.07 0.14 0.33 - -
30 mm - 0.1 0.2 0.5 - -
50 mm 0.1 0.2 0.3 0.8 - -
100 mm 0.1 0.4 0.7 1.7 - -
150 mm 0.1 0.5 1.0 2.5 12 24
250 mm - 0.9 1.8 4 20 40
c. Initial Column Efficiency Determination
1. Perform an efficiency test on the column before using it. Waters
recommends using a suitable solute mixture, as found in the
“Performance Test Chromatogram”, to analyze the column upon
receipt. However, if the column is used only for a single routine
assay, it may be more convenient to test the column under these
assay conditions. Keep a record of the initial column performance.
2. Determine the number of theoretical plates (N) and use this value for
periodic comparisons.
3. Repeat the test at predetermined intervals to track column
performance over time. Slight variations may be obtained on two
different HPLC systems due to the quality of the connections, operating
environment, system electronics, reagent quality, column condition
and operator technique.
[ Care and Use ManUal ]
Waters Spherisorb Columns 4
i i . column use
To ensure the continued high performance of Spherisorb columns,
follow these guidelines:
a. Guard Columns
Use a Waters Spherisorb guard column of matching chemistry and
particle size between the injector and main column. It is important to
use a matching guard column to protect the main column while not
compromising or changing the analytical resolution. Guard columns
need to be replaced at regular intervals as determined by sample
contamination. When system backpressure steadily increases above
a set pressure limit, it is usually an indication that the guard column
should be replaced. A sudden appearance of split peaks or other
changes in chromatographic performance is also indicative of a need
to replace the guard column.
b. Sample Preparation
1. Sample impurities often contribute to column contamination. One
option to avoid this is to use Oasis® solid-phase extraction cartridges/
columns or Sep-Pak® cartridges of the appropriate chemistry to clean
up the sample before analysis. Link to www.waters.com/sampleprep
2. It is preferable to prepare the sample in the operating mobile phase
or a mobile phase that is weaker (less organic modifier in the case
of reversed-phase chromatography, less polar modifier in the case
of normal-phase chromatography or hydrophilic-interaction chro-
matography, less salt in the case of ion exchange) than the mobile
phase for the best peak shape and sensitivity.
3. If the sample is not dissolved in the mobile phase, ensure that the
sample, solvent and mobile phases are miscible in order to avoid
sample and/or buffer precipitation. Filter sample with 0.2 μm
membranes to remove particulates. If the sample is dissolved in a
solvent that contains an organic modifier (e.g., acetonitrile, methanol,
etc.) ensure that the membrane material does not dissolve in the
solvent. Contact the membrane manufacturer with solvent compatibility
questions. Alternatively, centrifugation for 20 minutes at 8,000 rpm,
followed by the transfer of the supernatant liquid to an appropriate vial,
could be considered.
c. pH Range
The recommended operating pH range for Spherisorb columns is 2
to 8. A listing of commonly used buffers and additives is given in
Table 3. Additionally, the column lifetime will vary depending upon
the operating temperature, and the type and concentration of buffer
used. For example, the use of phosphate buffer at pH 8 in combina-
tion with elevated temperatures will lead to shorter column lifetimes.
[ Care and Use ManUal ]
Waters Spherisorb Columns 5
Table 3: Buffer Recommendations for Using Spherisorb Columns from pH 2 to 8
Additive or Buffer pKa
Buffer
Range
(± pH unit)
Volatility
Used for
Mass Spec?
Comments
Formic Acid 3.75 Volatile Yes
Maximum buffering obtained when used with ammonium formate salt. Used in
0.1-1.0% range.
Acetic Acid 4.76 Volatile Yes
Maximum buffering obtained when used with ammonium acetate salt. Used in
0.1-1.0% range.
Ammonmium Formate
(NH4COOH)
3.75 2.75-4.75 Volatile Yes
Used in the 1-10 mM range for LC/MS. Higher concentrations (typically 20 mM)
are recommended for UV applications.
Note: sodium or potassium salts are not volatile.
Trifluoroacetic Acid
(TFA)
0.3 Volatile Low Conc.
When used in LC/MS, due at signal suppression, it is generally recommended to
use TFA at coencentrations <0.1%.
Ammonium Acetate
(CH3COONH4)
4.76 3.76-5.76 Volatile Yes
Used in the 1-10 mM range for LC/MS. Higher concentrations (typically 20 mM)
are recommended for UV applications.
Note: sodium or potassium salts are not volatile.
Phosphate 1 2.15 1.15-3.15 Non-volatile No Traditional low pH buffer, good UV transparency.
Phosphate 2 7.2 6.20-8.20 Non-volatile No
Above pH 7, reduce temperature/concentration and use guard column to maxmize
lifetime.
d. Solvents
To maintain maximum column performance, use high quality
chromatography grade solvents. Filter all aqueous buffers prior
to use. Pall Gelman Laboratory Acrodisc® filters are recommended.
(Please refer to the filtration section of the Waters Chromatography
Columns and Supplies Catalog or the Waters web site (www.waters.com)
for additional product information.) Solvents containing suspended
particulate materials will generally clog the outside surface of
the inlet distribution frit of the column. This will result in higher
operating pressure and poorer performance. Degas all solvents
thoroughly before use to prevent bubble formation in the pump and
detector. The use of an on-line degassing unit is also recommended.
This is especially important when running low pressure gradients
since bubble formation can occur as a result of aqueous and organic
solvent mixing during the gradient.
e. Pressure
Spherisorb columns can tolerate pressures of up to 6,000 psi (400 bar
or 40 MPa) although long-term, routine operating pressures greater
than 4,000 – 5,000 psi should be avoided in order to maximize column
and system lifetimes.
f. Temperature
Temperatures between 20 – 45 ˚C are recommended for operating
Spherisorb columns in order to enhance selectivity, lower
solvent viscosity and increase mass transfer rates. However, any
temperature above ambient may have a negative effect on lifetime
which will vary depending on the pH and buffer conditions used.
i i i . scalinG up/doWn isoc rat ic met hods
The following formulas will allow scale up or scale down, while
maintaining the same linear velocity, and provide new sample
loading values:
If column i.d. and length are altered:
F2 = F1(r2/r1)2
or
Injection volume1 = Injection volume2 (r2/r1)2(L2/L1)
Where: r = radius of the column, in mm
F = flow rate, in mL/min
L = length of column, in mm
1 = original, or reference column
2 = new column
[ Care and Use ManUal ]
Waters Spherisorb Columns 6
iv. t roubleshoot inG
Changes in retention time, resolution, or backpressure are often due
to column contamination. See “Column Cleaning, Regenerating and
Storage”. Information on column troubleshooting problems may be
found in the current Waters Chromatography Columns and Supplies
Catalog. You can also download a copy of the HPLC Troubleshooting
Guide at www.waters.com, and in the Search Box, enter WA20769.
v. column cleaninG, reGenerat inG and storaGe
a. Cleaning and Regenerating
Changes in peak shape, peak splitting, shoulders on the peak, shifts in
retention, change in resolution or increasing backpressure may indicate
contamination of the column. Changing the guard column being used
will often restore column performance. If not (or if no guard column
is being used), follow the procedures detailed below. To prevent
potential contamination from affecting detector performance, it is
recommended that any detector(s) be disconnected from the effluent
flow of the column during cleaning. Reversing the direction of the
flow through the column (backflushing) may sometimes improve the
effectiveness of any cleaning procedure.
1. Reversed-Phase Columns
Flushing with a neat organic solvent, taking care not to precipitate
buffers, is usually sufficient to remove most contaminant. If the
flushing procedure does not solve the problem, purge the column
using the following cleaning and regeneration procedures. Use the
cleaning routine that matches the properties of the samples and/or
what you believe is contaminating the column (see Table 4). Flush
columns with 20-column volumes of HPLC-grade solvents (e.g.,
80 mL total for 4.6 x 250 mm column). Increasing mobile-phase
temperature to 35-55 ˚C increases cleaning efficiency. If the column
performance is poor after regenerating and cleaning, call your local
Waters office for additional support.
Table 4: Column Sequence or Options
Polar Samples Non-polar Samples Proteinaceous Samples
1. Water
1. Isopropanol (or an
appropriate isopropanol/
water mixture*)
Option 1: Inject repeated
aliquots of dimethyl
sulfoxide (DMSO)
2. Methanol 2. Tetrahydrofuran (THF) Option 2: gradient of
10-90% B where:
A= 0.1% trifluoroacetic
acid (TFA) in water
B= 0.1% trifluoroacetic acid
(TFA) in acetonitrile (CH3CN)
3. Tetrahydrofuran 3. Dichloromethane
4. Methanol 4. Hexane
5. Water
5. Isopropanol (followed by
an appropriate isopropanol/
water mixture*)
Option 3: Flush column with
7M guanidine hydrochloride
or 7M urea
6. Mobile Phase 6. Mobile Phase
* Use low organic solvent content to avoid precipitating buffers.
2. Normal-Phase Columns
To regenerate, pump 20-30 column volumes each of
dichloromethane and isopropanol through the column. Other wash
solvents such as tetrahydrofurane (THF) may also be selected based
on the suspected contamination.
Guard columns need to be replaced at regular intervals, as
determined by sample contamination. When system backpressure
steadily increases above a set pressure limit, it is usually an
indication that the guard column should be replaced. A sudden
appearance of split peaks is also indicative of a need to replace the
guard column.
b. Storage
Completely seal the column to avoid evaporation and drying out of
the bed.
1. Reversed-Phase Columns
For periods longer than four days at room temperature, store the
column (with the exception of cyano chemistry columns) in 100%
acetonitrile at room temperature. For elevated temperature applica-
tions, store immediately after use in 100% acetonitrile for the best
column lifetime. Do not store columns in buffered eluents. If the
mobile phase contained a buffer salt, flush the column with 10 col-
umn volumes of HPLC grade water (see Table 2 for common column
volumes) and replace with 100% acetonitrile for storage. Failure to
perform this intermediate step could result in precipitation of the
[ Care and Use ManUal ]
Waters Spherisorb Columns 7
buffer salt in the column when 100% acetonitrile is introduced.
Note: The exception of the storage recommendations above are cyano (CN)
columns used for reversed-phase methods (CN-RP). These columns should never
be stored in 100% intermediate polarity solvents (acetonitrile, methanol, IPA).
Store CN-RP columns in 0.1 M ammonium acetate/acetonitrile 50/50.
2. Normal-Phase Columns
For rapid equilibration upon start-up, store your normal-phase
column in the mobile phase that is commonly used. Completely seal
column to avoid evaporation and drying out of the bed.
v i. connect inG t he column to t he hplc
a. Column Connectors and System Tubing Considerations
All Spherisorb column and cartridges have Parker style endfittings.
Tools needed for Spherisorb analytical column:
1/2 inch wrench
9/16 inch wrench
Tools needed for Spherisorb analytical cartridge column:
1/4 inch wrench
Handle the column with care. Do not drop or hit the column on a hard
surface as it may disturb the bed and affect its performance.
1. Correct connection of 1/16 inch outer diameter stainless steel tubing
leading to and from the column is essential for high quality chromato-
graphic results.
2. When using standard stainless steel compression screw fittings, it is
important to ensure proper fit of the 1/16 inch outer diameter stainless
steel tubing. When tightening or loosening the compression screw,
place a 5/16 inch wrench on the compression screw and a 1/2 inch
wrench on the hex head of the column endfitting.
Note: If one of the wrenches is improperly placed on
本文档为【waters sax】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。