• 0 -1 ppm oxygen
• No sc rubbers
• No zero gas
• No ma in tenance
• No hass le
Teledyne’s Model 3000TA-XL is a precision, cost effective instrument specifically designed to measure trace oxygen
in the 0-1 ppm region. Drawing on the solid performance
of Teledyne’s highly successful 3000 series platform,
the 3000TA-XL provides the user with a wide variety of
performance features and options to satisfy virtually any
application.
FLEXIBILITY
The analyzer offers three dynamic, user-configurable ranges
and can be calibrated in any range. Two fully programmable
concentration alarms and relays provide the versatility to
satisfy nearly any requirement.
CONVENIENCE
The analyzer comes standard with two sets of analog outputs,
0-1 VDC and isolated 4-20 mADC for oxygen concentration
and range identification. Additionally, a bi-directional RS-232C
serial communication interface provides for remote monitoring
and control of span and zero functions.
MAINTENANCE FREE SENSOR
The heart of the 3000TA-XL analyzer is Teledyne’s micro-
fuel cell sensor that sets the industry standard for oxygen
analysis. The unit incorporates the B-2CXL sensor, which
is specifically designed for 0-1 ppm full scale oxygen
analysis applications.
The micro-fuel cell is specific to oxygen and can be used
in a variety of background gases including hydrocarbons.
With no electrolyte to change or electrodes to clean, it is
maintenance free.
APPLICATIONS
• Air liquefaction and separation
• Pure, gaseous hydrocarbon stream monitoring
• Semiconductor manufacturing
• Protective atmospheric blanketing of primary liquid
feedstock and flammable liquids
• Process monitoring of gaseous monomers
• Gas purity certification
ADVANTAGES
• 0-1 ppm full scale capability
• Linearity of analysis across three user-selectable ranges
• Remote access to auto span and auto zero functions
• Autoranging to follow process upsets
• Maintenance-free sensor
Built for Reliability and Performance
TELEDYNE
ANALYTICAL INSTRUMENTS
High Sensitivity Trace
Oxygen Analyzer
MODEL 3000TA-XL
High Sensitivity Trace
Oxygen Analyzer
Accuracy: ±100 ppb if the instrument is not zeroed
(at constant temperature)
±10 ppb if the instrument is zeroed (at
constant temperature)
Sensitivity: 0.5% of FS
Response time: 90% of FS at 77°F (25°C) 10 seconds
for 0-1000 ppm or higher ranges; 60
seconds for 0-10 ppm range 100 seconds
for 0-1 ppm range
Operating temp: 50° to 95°F (5° to 35°C)
Signal output: Analytical measurement 0-1 VDC and
isolated 4-20mADC
Range ID output: 0-1 VDC and isolated 4-20 mADC
Max. load impedance: 4-20 mA isolated output -- 1000 ohms
Analysis display: 5 digit red LED, 3/5” high numerals
Menu display: 20 character, 2 line vacuum fluorescent
Data lines: Bi-directional RS-232C serial interface,
baud rate 2400
Power requirements: Universal AC input ranges, 85 to
230VAC, 50 / 60 Hz
Max. power consumption: 20 VA
Oxygen sensor: Micro-fuel cell, Class B-2CXL
Sample gas inlet pressure: Min 2-3 psig
Wetted parts: 316 stainless steel
Sample inletconnection: 1/4” VCR
Area classification: General purpose
Mounting: Flush panel mount
Dimensions: 8.70” W x 6.96” H x 12.2” D (case)
10.79” W x 7.46” H (front panel)
Sample temperature: 0 to 50°C
Detector wetted parts: Glass, steel, gold, viton, acrylic glass,
epoxy resin
MODEL 3000TA-XL TRACE OXygEn AnALyzER
STANDARD FEATURES
• Class B-2CXL micro-fuel cell sensor
• Three user-selectable ranges plus cal range
• Signal output: 0-1 VDC & isolated 4-20 mADC
• Programmable autoranging
• Range ID output: 0-1 VDC & isolated 4-20 mADC
• Range ID contacts (quantity 4) Form A normally open contacts, 3A
resistive
• Two fully adjustable concentration alarm set points with
programmable relay functions, Form C contacts, 3A resistive
• Calibration contact span/zero, Form A normally open contact, 3A
resistive
• 12 VDC output to drive remotely operated calibration valves
(Customer supplied valves and 24 VDC signal to instrument)
• Self diagnostics with Form C failure alarm contacts
• Full duplex RS-232C communication link
• Five digit oxygen concentration LED display
• 2x20 alphanumeric vacuum fluorescent display for set up and
diagnostics
• 316 stainless steel sampling system
• Sample flow indicator and control valve
• Universal power supply: 85-230 VAC / 50-60 Hz
OPTIONS
-V Vacuum service
-K 19” rack mount with either one or two
units mounted side by side
SPECIFICATIONS
Ranges: 3 range customer selectable (Minimum
0-1 ppm FS)
Calibration range: Unit may be calibrated in any range up to
25%
© 2006 Teledyne Analytical Instruments, A Teledyne Technologies Company.
All rights reserved. Printed in the USA. 09/06LD
Warranty
Instrument is warranted for 1 year against defects in material
or workmanship
nOTE: Specifications and features will vary with application. The above are established
and validated during design, but are not to be construed as test criteria for every prod-
uct. All specifications and features are subject to change without notice.
TELEDYNE
ANALYTICAL INSTRUMENTS
A Teledyne Technologies Company
16830 Chestnut Street
City of Industry, California 91748, USA
TEL: 626-934-1500 or 888-789-8168
FAX: 626-934-1651 EMAIL: ask_tai@teledyne.com
www.teledyne-ai.com
本文档为【3000ta-xl】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。