QUALITY ASSURANCE DOCUMENT
QUALIFICATION OF EQUIPMENT
ANNEX 2: QUALIFICATION OF GC EQUIPMENT
Full document title and
reference
Qualification of Equipment
Annex 2: Qualification of GC Equipment
PA/PH/OMCL (06) 86 DEF
Document type Guideline
Legislative basis
Date of first adoption May 2006
Date of original entry
into force
June 2006
Date of entry into force
of revised document
October 2006
Previous titles/other
references
This document replaces part of document PA/PH/OMCL (06) 46
DEF
Custodian
Organisation
The present document was elaborated by the OMCL
Network/EDQM of the Council of Europe
Concerned Network GEON
ANNEX 2 OF THE OMCL NETWORK GUIDELINE
“QUALIFICATION OF EQUIPMENT”
QUALIFICATION OF GC EQUIPMENT
Introduction
The present document is the second Annex of the core document “Qualification of
Equipment”, and it should be used in combination with it when planning, performing and
documenting the GC equipment qualification process.
The core document contains the general introduction and the Level I and II of qualification,
common to all type of instruments, and the present annex contains GC instrument-related
recommendations on parameters to be checked and the corresponding typical acceptance
limits, as well as practical examples on the methodology that can be used to carry out these
checks.
The tests proposed in the Level III and IV of qualification are based on an overall approach,
in which several parameters are checked at the same time in a combined test procedure, to
obtain information on the overall system performance (e.g. peak area precision, retention time
precision, temperature programme reproducibility, etc).
Nevertheless, it should be noted that it is also acceptable to check these parameters
individually by using other well-defined procedures.
TABLE III
Level III. Periodic and motivated instrument checks
Examples of requirements for GC instruments with FID
Instrument
module
Parameter to be checked Typical tolerance limits
1.1 Injector leak test
Pressure drop ≤ 15 kPa
within 5 minutes
1.2. Pressure/flow accuracy and stability Covered by overall test 1
1.3. Repeatability of injection (overall test 1)
- In split mode
- In split less mode
RSD ≤ 3.0%
RSD ≤ 3.0%
1.4. Injector temperature accuracy and stability Covered by overall test 2
1. Inlet
system
1.5. Carry-over (overall test 3) ≤ 0.2%
2. Oven
2.1. Repeatability of oven temperature
characteristics
Covered by overall test 2
3.1. Linearity (overall test 3) r2 ≥ 0.999
3.2. Constant detector response Covered by overall test 1 or 2
3.3. Noise See Annex I
3. FID
detector
3.3. Drift See Annex I
TABLE IV
Level IV. In-use instrument checks
Examples of requirements for GC instruments with FID
Parameter to be checked Typical tolerance limits
1. System suitability check for the
method
According to Ph. Eur. or MAH dossier or
validated in-house method
2. Peak area precision
RSD ≤ 3.0% unless otherwise prescribed*
3. Retention time repeatability
RSD ≤ 2.0%
4. Sensitivity (where relevant, e.g. for
related substances tests)
According to Ph. Eur. or MAH dossier or
validated in-house method
* This is to be defined in conjunction with the target concentration of the analyte
All parameters given here should be checked when performing analyses under the working
conditions for the actual sample determinations. Normally, the test and reference solutions to
be prepared for this purpose are given as a part of the method.
ANNEX I
Level III. Periodic and motivated instrument checks
Practical examples of tests and their associated tolerance limits for several parameters related
to the performance of the different modules of a GC are presented below.
These examples can be considered by the OMCLs as possible approaches to perform the
Level III of the equipment qualification process: “Periodic and motivated instrument checks”.
Several tests are proposed to check various parameters at the same time (overall tests). In
order to run the tests in a more economical way, other suitable solutions can be used, as for
example, the “Grob Test” mixture, available from different suppliers (e.g. Alltech, Sigma,
Thames Restek). This commercial solution should be appropriate to the column material used.
It is recommended to run the overall tests by using always the same test column, exclusively
dedicated to qualification purposes, to guarantee reproducible conditions.
1. INLET SYSTEM
The following tests are proposed for the periodic and motivated check of the GC Inlet System.
1.1. INJECTOR LEAK TEST
Method:
If not otherwise specified by the instrument manufacturer, the leak test is carried out
according to the procedure laid down in the instrument manual or by the built in automatic
leak check procedure of the instrument.
Otherwise use the test described below:
Disconnect the column from the injector and close the injector outlet with a sealed cap.
Close the septum purge and the bypass.
Adjust the flow and pressure controller to the maximal possible value of the pressure gauge.
Adjust the flow controller to zero.
Read the pressure after 1 minute and record the value.
Record the pressure after 5 minutes.
Limits:
Pressure drop ≤ 15 kPa within 5 minutes.
1.2. INLET PRESSURE/FLOW ACCURACY AND STABILITY
A direct measurement of these parameters was not deemed practical or necessary, but the
optimal conditions of flow/pressure can be verified by the overall test 1.
Limits: refer to overall test 1.
1.3. REPEATABILITY OF INJECTION
The verification of this parameter is covered by the overall test 1.
This test is to be performed in both split and split less mode.
Limits: refer to overall test 1.
1.4. INJECTOR TEMPERATURE ACCURACY AND STABILITY
Due to the fact that the temperature cannot be reliably measured without opening and
modifying the system and due to the difficulties of introducing a probe inside this module, the
verification of this parameter is considered to be covered by the overall test 2.
Limits: refer to overall test 2.
1.5. INJECTOR CARRY OVER
After having injected the solutions for the linearity test of the FID detector, in increasing
order, inject the blank and measure the peaks that correspond to the major peaks (= analytes)
in the linearity solutions.
The verification of this parameter is covered by the overall test 3.
Limits: refer to overall test 3.
2. OVEN
2.1. REPEATABILITY OF THE OVEN TEMPERATURE CHARACTERISTICS
Due to the fact that the temperature cannot be reliably measured without opening and
modifying the system conditions and that even when introducing a probe inside the oven, its
location would not reflect the real temperature conditions at all points, the verification of this
parameter is covered by the overall tests 2A and 2B.
Limits: refer to overall test 2.
3. FID DETECTOR
The following tests are proposed for the periodic and motivated check of the GC FID
detector.
3.1. FID DETECTOR LINEARITY
Increasing amounts of analyte are injected and a linear response should be obtained.
The verification of this parameter is covered by the overall test 3.
Limits: refer to overall test 3.
3.2. CONSTANT FID DETECTOR RESPONSE
The proper and reproducible functioning of the FID can be demonstrated by checking the
peak areas obtained from a pre-defined standard solution.
The verification of this parameter is covered by the overall test 1 or 2.
Limits: refer to overall test 1 or 2.
3.3. FID DETECTOR NOISE AND DRIFT
If the instrument has a built-in automatic system for the verification of the noise and drift,
follow the manufacturer’s instructions and apply the defined acceptance criteria. Otherwise,
use the test described below:
Settings:
Column installed
Suitable flow, depending on column length/diameter
No injection
Oven temperature: 40°C
Detector on and heated at working temperature (270-300°C)
Method:
After stabilisation of the system, record the signal for 15 minutes.
Noise: evaluate 10 periods of 1 minute and calculate the mean value.
Drift: evaluate the slope of the baseline over the 15 minutes.
Limits:
The acceptance criteria for these parameters have to be chosen in accordance with the
instrument vendor’s instructions and the intended use of the instrument. If no instructions are
given, the user has to pre-define these acceptance criteria by taking into account the previous
experience and the intended use of the instrument.
No fixed values can be pre-defined in this guideline due to the high variety of integration
systems used and consequently the acceptance criteria may be expressed in different units
(voltage, current, arbitrary units per time).
OVERALL TEST 1
The overall test 1 covers the following parameters:
- Pressure/flow accuracy and stability in the inlet system: Retention time repeatability
- Repeatability of injection: peak area precision
- In split mode
- In split less mode
The test may be combined with overall test 3.
Split mode:
Test solution:
1-octanol in n-hexane 1% (v/v).
Settings:
Column: SPB-1 (30m x 0.32mm ID x 0.25µm film)
Carrier gas: He
Velocity: 25cm/sec
Split: 1:100
Injection: 1µl
Injector temperature: 220°C
Oven temperature: 100°C isotherm
Detector temperature: 300°C
Runtime: 8 min
Retention time of 1-octanol: about 5 min
Split less mode:
Stock solution: 1-octanol in n-hexane 1% (v/v)
Test solution: Dilute 10 ml of the stock solution with n-hexane to 100 ml (corresponds to
1µl/ml of 1-octanol in n-hexane)
Settings:
Column: SPB-1, 30m, 0.32mm ID, 0.25µm film
Carrier: He
Velocity: 30cm/sec
Split less injection: purge valve closed during 2 min
Injection: 0.2µl of the test solution
Injector Temperature: 220°C
Oven Temperature: Initial 60°C for 4 min, 15°C/min. up to 135°C, final time 1min
Detector temperature: 300°C
Runtime: 9.5 min
Retention time of 1-octanol: about 8 min
Method:
Carry out 6 consecutive injections of the test solution and calculate the RSD of the different
peak areas and retention times.
Limits:
Retention time repeatability: the RSD of the retention times should be ≤ 2.0%
Peak area precision (split and split less mode): the RSD of the peak areas should be ≤ 3.0%
OVERALL TEST 2
The overall test 2 covers the following parameters:
- Injector, oven and detector temperature accuracy and stability: retention time
repeatability
Two alternative tests are proposed:
Overall test 2A
Test solution:
0.035 ml 1-octanol
0.035 ml 2-octanone
0.035 ml 2,6-dimethylanilin
0.035 ml n-tridecane
0.035 ml n-tetradecane
35 mg n-eicosane
dissolved in 50 ml Dichloromethane
Settings:
Column: SPB-1 (30m x 0.32mm ID x 0.25µm film)
Carrier gas: Helium
Velocity: 25 cm/s
Split: 1:100
Injection volume: 1 µl
Injector temperature: 220°C
Detector: FID
Detector temperature: 300°C
Gradient programme: 60°C (4 min), 5°C/min, 270°C (3 min)
Method:
Inject the solution twice and calculate the relative retention times in relation to n-eicosane
(RRT = 1)
The following table shows the approximately expected relative retention times.
Analyte 1-octanol 2-octanone 2,6-dimethylaniline n-tridecane n-tetradecane
RRT 0.30 0.22 0.37 0.52 0.60
Limits:
The RSD of each RRT from two consecutive injections should be ≤ 1.0%
Overall test 2B
Test Solution:
1.0% (W/W) n-Nonane and Hexadecane in Tetradecane.
Settings:
Column: Ultra-1 (25m x 0.32mm ID x 0.52µm film)
Injection volume: 1 µl
Solvent: Tetradecane
Oven temperature: 110°C
Gradient programme: 110°C, 20°C/min, 180°C (final time: 3.5 min)
Detector temperature: 250°C
Injector temperature: 200°C
Detector: FID
Flow rates: Carrier gas (Helium): 2 ± 0.2 ml/min
Hydrogen: 30 ± 1.0 ml/min
Air: 400 ± 20.0 ml/min
Makeup (Nitrogen): 28 ± 1.0 ml/min
Split ratio: 15
Split vent: 30 ± 3.0 ml/min
Septum purge: 3-5 ml/min
Method:
Allow the system to equilibrate.
Injection sequence:
1) blank (Tetradecane)
2) 6 replicates of the test solution. Calculate the mean of the retention times and peak
areas and the relative standard deviation of n-Nonane and n-Hexadecane.
Limits:
Retention time repeatability: RSD of the peak retention times of the 6 replicates ≤ 2%
Retention time (Rt) accuracy: for this example, the retention time ranges shown in the table
below are proposed. Nevertheless, individual ranges should be predefined by the laboratory
depending on the column used (e.g. Rt ± 0.2 min).
Compound Rt (min)
n-Nonane (C9) 1.3 – 1.7
Tetradecane (C14) 4.0 – 4.7
Hexadecane (C16) 5.1 – 6.0
OVERALL TEST 3
This test is a modified version of the overall test 1 to be used for the verification of:
- Detector linearity: linearity of the areas recorded
- Injector carry-over: area recorded in the blank run
It is described for both split and split less mode and may be combined with overall test 1.
Split mode:
Test solution: 1-octanol in n-hexane 1% (v/v)
Prepare further reference solutions by diluting the test solution as described below.
Settings: see overall test 1
Injection sequence:
5.0 ml of the test solution diluted to 25.0 ml with n-hexane (2 µl/ml): 2 injections
10.0 ml of the test solution diluted to 25.0 ml with n-hexane (4 µl/ml): 2 injections
15.0 ml of the test solution diluted to 25.0 ml with n-hexane (6 µl/ml): 2 injections
20.0 ml of the test solution diluted to 25.0 ml with n-hexane (8 µl/ml): 2 injections
if combined with overall test 1 for repeatability: test solution (10 µl/ml): 6 injections
n-hexane as blank (carry over)
Split less mode:
Stock solution: 1-octanol in n-hexane 1% (v/v)
Test solution: Dilute 10 ml of the stock solution with n-hexane to 100 ml (corresponds to
1µl/ml of 1-octanol in n-hexane).
Prepare further reference solutions by diluting the test solution with n-hexane.
Settings: see overall test 1
Injection sequence:
5.0 ml of the test solution diluted to 25.0 ml with n-hexane (0.2 µl/ml): 2 injections
10.0 ml of the test solution diluted to 25.0 ml with n-hexane (0.4 µl/ml): 2 injections
15.0 ml of the test solution diluted to 25.0 ml with n-hexane (0.6 µl/ml): 2 injections
20.0 ml of the test solution diluted to 25.0 ml with n-hexane (0.8 µl/ml): 2 injections
if combined with overall test 1 for repeatability: test solution (1 µl/ml): 6 injections
n-hexane as blank (carry over)
Limits:
Linearity: coefficient of correlation of the calibration line obtained with the reference
solutions and the test solution: r2 ≥ 0.999.
Carry-over: the percentage of the peak area corresponding to the analyte in the blank solution
should be ≤ 0.2% of the peak area of this analyte in the chromatogram obtained with the
solution with the highest concentration within the sequence.
本文档为【annex_2_qualification_of_gc_equipment】,请使用软件OFFICE或WPS软件打开。作品中的文字与图均可以修改和编辑,
图片更改请在作品中右键图片并更换,文字修改请直接点击文字进行修改,也可以新增和删除文档中的内容。
该文档来自用户分享,如有侵权行为请发邮件ishare@vip.sina.com联系网站客服,我们会及时删除。
[版权声明] 本站所有资料为用户分享产生,若发现您的权利被侵害,请联系客服邮件isharekefu@iask.cn,我们尽快处理。
本作品所展示的图片、画像、字体、音乐的版权可能需版权方额外授权,请谨慎使用。
网站提供的党政主题相关内容(国旗、国徽、党徽..)目的在于配合国家政策宣传,仅限个人学习分享使用,禁止用于任何广告和商用目的。